Normal Circulation

The cardiovascular system has ten unique characteristics that make it an unusually complicated hydraulic system. Understanding how the cardiovascular system functions requires insight into a larger set of variables than that which governs the function of most pump, pipe, and fluid systems found in the world of man-made machines. The ten unique characteristics peculiar to the cardiovascular system are:
  1. The system is a closed circle rather than being open-ended and linear.

     

  2. The system is elastic rather than rigid.

     

  3. The system is filled with liquid at a positive mean pressure ("mean cardiovascular pressure"), which exists independent of the pumping action of the heart.

     

  4. The right and left ventricles, which pump into the same system that they pump out of, are in series with two interposed vascular beds (systemic and pulmonary).

     

  5. The heart fills passively, rather than by actively sucking.

     

  6. As a consequence of the heart's passive filling, the circulation rate is normally regulated by peripheral-vascular factors, rather than by cardiac variables.

     

  7. The flow from the heart is intermittent, while the flow to it is continuous.

     

  8. Normally, there is an excess expenditure of energy by the heart needed for the circulation rate imposed by peripheral vascular regulators ("pump energy excess").

     

  9. Normally, ventricular capacity is in excess of the diastolic filling volume ("pump capacity excess").

     

  10. The slowing effect of any vascular resistance on flow rate depends on its location, with reference to upstream compliance, as well as its magnitude.

     

In order to understand circulatory phenomena in the elastic circular hydraulic system (Fig. 1), where every point is both upstream and downstream from every other point, and where the non-sucking ventricles pump out of the same system they pump into, we need to examine the ten unique factors individually, before we can amalgamate them into a meaningful whole.

 

VENTRICLES (THE PUMPS)

To clarify and facilitate understanding of the features peculiar to the heart, it is helpful to compare three major types of pumps. The heart�s unique characteristics as a pump are of paramount importance in understanding how the cardiovascular system works.

PUMP TYPE #1: This type of pump both sucks and forcibly ejects fluid. This pump uses energy both to actively fill at its inlet and to empty its contents at its outlet. Two examples of this type of pump are: (1) a piston pump, which expends energy to suck in the stroke volume which is then forcibly ejected; and (2) a roller pump, which sucks at its inlet by the recoil of the resilient tubing that has been compressed by the roller as it moves forward, ejecting the fluid in front of it. With type #1 pumps, the output in a hydraulic system is determined exclusively by two pump variables: the stroke rate and the stroke volume. The reason for discussing this is that the heart is a different type of pump, and those two variables are frequently and erroneously projected onto cardiovascular function, in a way in which they do not apply.

PUMP TYPE #2: This pump sucks and blows but, instead of producing a specific flow rate, creates a specific pressure gradient between its inlet and outlet. Centrifugal pumps fall into this category. With this type, two pump factors (rate and power), as well as two non-pump factors (pressure and resistance in the system), effect output. As the heart is not such a pump, there is danger in borrowing explanations of cardiovascular function from hydraulic systems containing centrifugal pumps.

PUMP TYPE #3: This type of pump is passive filling, and does not suck at its inlet. It expends no energy to fill, it only expends energy to empty. An example of this type of pump is the urinary bladder. It is a flaccid, hollow organ that does not create any negative pressure or suck on the ureters or kidneys to fill. The bladder merely exerts energy to empty. To calculate the flow of urine for any given period, you can obtain the answer by multiplying the stroke rate times the stroke volume. However, it is important to underline here that those two things, stroke rate and volume, are not determinants of bladder output. You cannot increase the output merely by changing the rate of bladder contraction (stroke rate). The urinary bladder cannot expend energy to increase its filling and thus its stroke volume. This type of pump, even though it does the work and thus produces the flow, is totally dependent upon external factors (e.g., renal function) to determine the output. At a given rate of urinary production, stroke rate and stroke volume are reciprocals of one another. If the bladder is emptied twice as often, the stroke volume will be one half as much.

THE RIGHT AND LEFT VENTRICLES ARE TWO TYPE #3 PUMPS:

The heart, like the urinary bladder, is a hollow muscular organ that does not suck to fill, but produces circulation by ejecting whatever fluid enters it at diastole. During normal function, the heart not only doesn't develop a pressure negative to the intrathoracic pressure, but it offers an impediment to filling because of its limited volume-pressure compliance.

The evidence that the heart fills passively, and does not suck to fill, is found in the data from innumerable heart catheterizations, which all show a positive diastolic pressure in the ventricles (Fig. 2, a to c). In fact, as ventricles fill, they not only do not suck, but they offer an increasing impediment to filling, as noted by the progressive increase in pressure toward the end of diastole (Fig. 2, c). Negative ventricular pressure, in relation to intrathoracic pressure, has not been found in physiologic states. The heart, like all other muscles in the body, expends energy and does work by contracting. It cannot expend energy to do mechanical work by forcibly elongating its fibers to suck like the type #1 pump, which uses energy to suck in a stroke volume. Do not confuse the negative pressure created in the chest by inspiration as negative heart pressure (Fig. 2, b). The chest can suck, the heart cannot.

One can calculate cardiac output in the same way as in the urinary bladder example: by multiplying stroke rate times stroke volume. Also, just as in that example, while stroke rate multiplied by stroke volume measures the amount of output, those variables are not determinants of that output. The heart, by filling passively, pumps out blood at a rate determined by the rate of blood coming to it. Given that the heart is a pump that produces the flow but has a flow rate determined by extra-cardiac factors, let us examine those factors that determine circulation rate.

 

DETERMINATION OF CARDIAC OUTPUT IN A CIRCULAR ELASTIC SYSTEM WITH PASSIVE FILLING PUMPS

With pumps that cannot suck to fill, there must be a positive pressure at the inlets for any blood to run into the ventricles, in order for there to be any pump output. If there is no pressure in the cardiovascular system, no blood can run into the ventricles and there can be no flow. Normally, there is a mean cardiovascular pressure above zero, which the heart distributes. The heart, rather than being responsible for the pressure in the vascular system, is a circulating device. It lowers the pressure at the ventricular inlets and raises it at the ventricular outlets. With a positive pressure in the cardiovascular system, when blood is ejected into the arterial side of the circle, a pressure gradient is created between the arteries and veins. This gradient causes blood to flow around the circle back to the ventricular inlets. Therefore, the output rate varies directly with the magnitude of that mean cardiovascular pressure. The higher the pressure, the higher the gradient, the greater the flow rate. The circular system being elastic, and having resistance and other impediments to flow, the energy from ventricular contraction does not transfer instantaneously around the circle after each heart beat, as would occur in a rigid system. The energy of venous flow is several heart beats behind that of ventricular ejection. The greater the elasticity and impediments to flow, the slower the flow rate.

Therefore, during normal function, cardiac output varies directly with the mean cardiovascular pressure and inversely with the impedance to blood flow to the heart.

I. MEAN CARDIOVASCULAR PRESSURE

Definition: The mean cardiovascular pressure is the pressure in the cardiovascular system with the circulation stopped, after the pressure has equalized between the arteries, capillaries, veins, and cardiac chambers. Do not confuse this pressure with central venous pressure, venous filling pressure, or mean-arterial pressure. Mean cardiovascular pressure is the pressure related to the blood volume and the compliance of the entire elastic cardiovascular compartment.

Measurement: Mean cardiovascular pressure is expressed in centimeters of water above ambient pressure, with zero being at mid-heart level. The mean cardiovascular pressure can be approximated during cardiac arrest. After arrest, a pressure equalization occurs between the various cardiovascular compartments in about thirty seconds. The arterial pressure falls and the venous pressure rises as some of the arterial blood moves into the veins during pressure equalization. Therefore, the mean cardiovascular pressure is always above venous pressure and below arterial pressure. Normally, mean cardiovascular pressure is between 15 and 18 cm. of water above mid-heart level. We have some approximation of its magnitude from fortuitous records of the arterial and venous pressures equalizing, obtained during short periods of cardiac arrest of patients in coronary care settings, in emergency rooms, or in operating rooms during heart surgery. Even in these situations, the resulting pressure can be regarded as only an approximation, as the shift in fluid, from hypoxia caused by lack of circulation, may have altered the vascular compliance. (See Appendix for clinical measurement technique)

Significance: Without mean cardiovascular pressure there would be no circulation. The heart doesn't generate the pressure in the vascular system, it merely distributes the mean cardiovascular pressure. The cardiac ventricles take the mean cardiovascular pressure and distribute it by raising the pressure on the arterial sides while lowering it on the venous sides. The two ventricles, being passively filling pumps, cannot suck, so they lower the inlet pressures toward — but never below — zero, in relation to the ambient pressure in the chest.

The higher the mean cardiovascular pressure, if the ventricles are not failing, the higher the ventricles can elevate the arterial pressure while reducing the venous pressure toward zero, and thus the greater the cardiac output. Conversely, the lower the mean cardiovascular pressure, the less the heart can raise the arterial pressure by lowering the venous pressure, and thus the lower the cardiac output.

Origin of the mean cardiovascular pressure: The mean cardiovascular pressure results from the volume of blood and the compliance of the cardiovascular system. The volume in the cardiovascular system results from an equilibrium between the rate of water, electrolytes, and other blood constituents entering the body by way of the gastrointestinal tract, and leaving the body primarily by the kidneys (Fig. 3). The mean cardiovascular pressure is the result of a continuing dynamic process.

 

MEAN CARDIOVASCULAR PRESSURE =
energy forcing fluid into the body / resistance to fluid loss from the body

Homeostatic maintenance of normal mean cardiovascular pressure:

(1) Slow feedback mechanism:

A slow homeostatic feedback mechanism tends to keep the mean cardiovascular pressure at a constant level: Elevation of the mean cardiovascular pressure above normal —› causes increase in cardiac output —› causes increased renal blood flow —› results in increased renal output —› thereby lowering blood volume and mean cardiovascular pressure back to normal. Conversely, low mean cardiovascular pressure —› low cardiac output —› low renal blood flow —› decreased renal output until the mean cardiovascular pressure is restored to normal by continuing fluid intake. With elevated mean cardiovascular pressure, the rate of return to normal is dependent on renal function, whereas, with low mean cardiovascular pressure the rate of return can vary greatly, depending on the rate of restoration of blood volume.

Clinical evidence of the homeostatic mechanism:

 

  • Response to weightlessness by going into orbit: In the absence of gravity, normal blood volume shifts centrally from the lower part of the body, thereby, increasing the mean cardiovascular pressure at heart level —› resulting in increased cardiac output —› causing greater renal blood flow —› leading to greater urinary output —› causing a decrease in blood volume and, thus, a decrease in mean cardiovascular pressure back to normal. The converse is found when astronauts return to gravity. It takes a few hours to restore normal mean cardiovascular pressure by intake of fluid and electrolytes after returning to earth, during which transition time they conserve the fluid they take in by putting out very little urine.

     

  • During any hypovolemic shock state, such as massive hemorrhage or severe dehydration, the urinary output goes abruptly down and remains low until restoration of normal mean cardiovascular pressure.

(2) Rapid mean cardiovascular pressure buffer mechanisms:

(a) Elasticity: The elasticity of the vascular system prevents sudden blood volume loss or gain from causing a linear, temporary change in mean cardiovascular pressure. Elasticity has, of course, an instantaneous buffer effect. Evidence of this is found in one's ability to give a pint of blood at the blood bank without going into severe low cardiac output. This buffer effect bolsters circulation while the blood volume — and, thus, mean cardiovascular pressure — is restored by subsequent oral intake of fluid.

(b) Vascular/extravascular equilibrium: There is a pressure equilibrium between the various extravascular compartments of the body and the cardiovascular space (Fig. 3). Changes in mean cardiovascular pressure result in shifts of fluid back and forth which tend to buffer sudden changes. This buffer system is fairly rapid but not instantaneous, as evidenced by the observation that a person going into severe shock from sudden loss of blood would not have had the same severe shock state if the loss had occurred more gradually over a period of an hour or so.

(3) Humeral and neuro-muscular-vascular reflexes:

These responses from stimuli, which alter vascular compliance, act as buffer systems. They prevent sudden changes in mean cardiovascular pressure from sudden position changes, such as going from lying to standing. They also buffer the effect of sudden loss of blood volume from hemorrhage.

II. IMPEDANCE TO THE FLOW OF BLOOD FROM THE OUTLETS AROUND TO THE INLETS OF THE VENTRICLES

Four factors tend to impede the flow of blood in the cardiovascular circle. Therefore, they are inverse determinants of cardiac output: (1) resistance, (2) elasticity, (3) limited compliance of the ventricles to filling, and (4) inertia of intermittent blood flow to the ventricles. The interrelationship of these four factors makes flow determination much more complicated than plain resistance, which is the single impediment in rigid, open ended, linear hydraulic systems.

(1) Resistance and (2) Elasticity

Unlike rigid linear hydraulic systems, where a given resistance may have the same effect on flow, irrespective of its location, the elasticity of the vascular system makes location of the resistance a significant parameter. Because of the elasticity of the circle, the location of any particular resistance determines to what extent that resistance has on impeding blood flow. A given resistance may have little or no effect in determining cardiac output if it is near the outlet of the ventricles, yet the same magnitude of resistance may have tremendous slowing effect on circulation if located near the inlets of the heart. Resistance points that have little compliant vascular bed "upstream" (arteriolar resistance), increase pump work but may not affect cardiac output significantly. The heart, except during failure, exerts enough energy to force the blood past any resistance near its outlet, with no hold up in flow. On the other hand, resistance located with a large compliant bed upstream has a tremendous effect on flow rate by slowing blood return to the ventricles. Venous sided resistance, with the compliance of the whole vascular bed upstream, is a major inhibitor of circulation rate. Thus, venous sided resistance is a major determinant of cardiac output, but arterial resistance is not. Interpolation of the additive effect of all resistance points in the circle on cardiac output must include the amount of compliant bed upstream from each resistance point.

All resistance factors — including blood viscosity, cross-section area of any vascular bed, margination of blood constituents, etc. — play roles in flow rate determination only when linked with their location to compliance upstream. Adding venous sided resistance of a magnitude that results in only a few mm. water pressure gradient may cause a marked reduction of flow. Whereas, increasing arteriolar resistance to the point of severe arterial hypertension may not appreciably change cardiac output.

In human cardiovascular physiology, the significance of resistance can only be understood when coupled with upstream compliance. It is not just the magnitude of resistance but the location of that resistance that determines its effect on circulation rate.

(3) Impediment to Ventricular Filling

The end-diastolic pressure: Ventricles not only do not suck to fill, they offer an impediment to filling. The left ventricle has thicker and stiffer walls than the right, so it tends to retard filling more than the right ventricle. Catheterization data shows end-diastolic pressure of five to ten centimeters of water above intrathoracic pressure. Don't confuse the negative pressure of inspiration, transmitted to the heart, as the heart sucking. The intracardiac pressure is always above that intrathoracic pressure (Fig. 2 at C).

(4) Inertia of Intermittent Blood Flow Offset by "The Atrial Effect"

Atrial function facilitates circulation by preventing the retarding effect that would otherwise occur from the intermittent inflow to the intermittent outflow ventricles. By being partially empty and distensible, atria prevent the interruption of venous flow to the heart that would occur during ventricular systole if the veins ended at the inlet valves of the heart.

Atria have four essential characteristics that cause them to promote continuous venous flow. (1) There are no atrial inlet valves to interrupt blood flow during atrial systole. (2) The atrial systole contractions are incomplete and thus do not contract to the extent that would block flow from the veins through the atria into the ventricles. During atrial systole, blood not only empties from the atria to the ventricles, but blood continues to flow uninterrupted from the veins right through the atria into the ventricles. (3) The atrial contractions must be gentle enough so that the force of contraction does not exert significant back pressure that would impede venous flow. (4) The "let go" of the atria must be timed so that they relax before the start of ventricular contraction, to be able to accept venous flow without interruption.

By preventing the inertia of interrupted venous flow that would otherwise occur at each ventricular systole, atria allow approximately 75% more cardiac output than would otherwise occur. The fact that atrial contraction is 15% of the amount of the succeeding ventricular ejection has led to the false conclusion that atria have their benefit by pumping up the ventricles (the so-called "atrial kick"). The real benefit is in preventing inertia and allowing uninterrupted venous flow.

The 20% to 25% increase in cardiac output from synchronized atrial function over that of atrial fibrillation doesn't belie the 75% contribution of the atrial effect, as atrial fibrillation eliminates only part of that effect. Atrial compliance, elasticity, and gravity help in emptying the atria at ventricular diastole during atrial fibrillation. Also, cardiac output during atrial dysfunction is buffered: the initial fall in circulation rate during atrial fibrillation reduces renal flow, thereby causing retention of water and the subsequent rise in mean cardiovascular pressure, which then partially offsets the slowing effect on circulation.

Thus, four factors impede the flow of blood around the cardiovascular circle: (1) resistance with (2) upstream compliance, (3) ventricular non-compliance, and (4) inertia if there is intermittent venous flow. The combined effect of these factors that impede the flow of blood to the inlet of the ventricles and, therefore, determine cardiac output in a negative way, will be referred to as: inlet impedance.

 

In summary, during normal physiology:

CARDIAC OUTPUT=FUNCTION OF MEAN CARDIOVASCULAR PRESSURE / INLET IMPEDANCE

 

AUTOMATIC BALANCING OF THE PULMONARY AND SYSTEMIC BLOOD VOLUMES

The passive filling characteristic of the ventricles is the feature that accounts for the automatic maintenance of blood volume equilibrium between the pulmonary and systemic vascular beds. With passively filled pumps, the relative blood volume in the two circuits is determined by their relative size and elasticity. On the other hand, if the ventricles were type #1 pumps, which actively fill, any discrepancy in the output of the two pumps, or flow through normal physiological shunts between the two vascular systems, would very quickly shift all of the blood volume into one circuit at the expense of the other, resulting in disaster.

The right and left ventricles, by filling passively, pump out whatever amount comes to them, determined by extracardiac factors. The blood volume equilibrium, therefore, is determined by the relative size and compliance of the two circuits and, to a minor extent, by the relative impediment of flow to the two pumps. It has been noted that the output of the two sides of the heart is never equal because there are physiological shunts, which connect one vascular bed to the other. The largest of these shunts, in normal physiology, are the bronchial arteries, which go from the systemic circuit to the lungs. The bronchial blood flow is a left-to-right shunt that accounts for the left ventricular output normally being at least 10% larger than the right. Because of the shunt, more blood returns to the left ventricle than the right, the left ventricle passively fills more than the right, thereby causing it to produce a greater output and thus the equilibrium is maintained. With other types of pumps this would not occur.

A dramatic illustration of this volume equilibrium, automatically being maintained during a massive discrepancy in output of the two pumps, is seen in atrial septal defects. In this case, the right ventricular output may be four or five times that of the left. Yet, the volume equilibrium is maintained. A large atrial septal defect virtually results in a single atrium above the two ventricles. The shunt occurs during ventricular diastole, because the right, thin walled ventricle is more distensible than the non-compliant thicker walled left ventricle. The blood in the common atrium goes the way of least resistance. The greater filling into the more compliant right ventricle results in greater right ventricular output. The greater right output goes to the lungs and then directly back to the right ventricle, returning again through the septal defect. Because of the passive filling, this results in no progressive increase in the blood volume in the lungs, and no disturbance in the maintenance of the blood volume equilibrium. After closure of the septal defect, resulting in much smaller right heart output, the volume equilibrium remains. This equilibrium, which persists after such sudden, massive changes in right heart output, occurs automatically because of the passive filling characteristic of the ventricles. It is the physical characteristics of the two vascular beds (e.g., relative size, compliance, and impediment to flow), that determines the volume balance with passive filling ventricles.

 

The passive filling of the ventricles accounts for the maintenance of blood volume equilibrium between the systemic and pulmonary vascular systems.
HEART RATE AND STROKE VOLUME

Normally, the passively filling ventricles are not maximally filled at diastole. Also, they are exerting excess energy over that needed to eject blood at the flow rate of blood entering the ventricles. With such volume and energy reserve, if blood enters the heart faster, the flow rate can go up without any change in strength of contraction or heart rate, unless the heart is being filled maximally (heart failure). Conversely, if blood enters the ventricles at a slower flow rate, the output will go down irrespective of any lowering of the rate or decrease in strength of cardiac contraction.

Thus, at a given flow rate, with the normal excess pump power, in both strength and rate, the heart rate and stroke volume become reciprocals of one another, as long as the heart doesn�t go into failure. At a given flow rate, decreasing the heart rate increases the stroke volume; while increasing the heart rate decreases the stroke volume. The observation — that cardiac output usually parallels the cardiac rate and strength of contraction — has resulted in the fallacious conclusion, based on post hoc ergo propter hoc reasoning ("following this, therefore because of this"), that they are cause and effect.

Even though these two variables do not normally control cardiac output, their variability has physiological benefit in energy conservation. During low circulation rates, when the heart is not filling to its maximal volume during diastole, the heart down shifts its rate and strength of contraction, thus conserving energy. Also, during high output states, it increases its rate and strength of contraction, thus preventing failure from any limitation of output that might be imposed by ventricular chamber size. If the heart functioned at its maximum strength of contraction and a rate of 150 beats a minute, the cardiac output would go up and down just as it does normally. But what a waste of energy would result! So the variability of rate and strength of contraction have only ecological value: that of conserving world food supply.

Two mechanisms cause the heart to roughly parallel its energy expenditure with cardiac output, preventing failure during high output states and saving energy during low output: (1) The neural and humeral stimuli that increase mean cardiovascular pressure are also those that increase heart rate and strength of contraction. (2) Increased stretching of the ventricles at diastole causes some increased strength of contraction at systole (Starling's law of the heart). While this observation of Starling is true and contributes to the paralleling of cardiac output to strength of contraction, it is probably not a major determinant of energy expenditure, as strong cardiac contractions continue even when the heart is completely empty, as seen during cardiac bypass surgery.

The reciprocal relationship of heart rate and stroke volume can be demonstrated in patients with complete heart block by varying the rate of firing of their pacemakers. Variations in rate, within physiological levels, above that required to prevent complete diastolic ventricular filling, are associated with no change in cardiac output. Therefore, a compromise rate of 78 is a commonly used setting that allows a wide range of activity acceptable to most patients.

 

Conclusion: During normal physiologic states, there is always pump energy excess over that used in circulation.
EXERCISE AND CARDIAC OUTPUT INCREASE

Two factors cause the increase in cardiac output during exercise: Exercise causes the cardiovascular impedance to be decreased and the mean cardiovascular pressure to be increased. The intermittent skeletal muscle contractions around venous beds, which contain one-way valves, act as a peripheral pump which overcomes significant impedance to flow. Neuro-humeral reflexes speed the heart rate, which slightly lowers the inlet impedance to the heart, by lowering the end-diastolic pressure over what it would otherwise be. The increased heart rate guarantees excess energy expenditure, thus preventing cardiac power failure at the higher circulation rate. The neuro-humeral response to exercise also throws the vascular system into spasm, thereby increasing the mean cardiovascular pressure from the "G-suit" effect of tensing the body in general. These are temporary changes, lasting only during the exercise period, which do not effect the long term homeostatic control of circulation.

DISTRIBUTION OF BLOOD FLOW: PULSATILE BLOOD FLOW AND ARTERIOLAR RESISTANCE

Pulsatile arterial blood flow tends to result in diffuse, fairly equal distribution of blood to all tissues of the body. This phenomenon would not occur with a non-pulsatile steady flow. Arteriolar resistance variability from time to time and from place to place, superimposed on the otherwise diffuse distribution, controls preferential blood flow to specific areas, with physiologic benefit. The diverting of a greater portion of cardiac output to the digestive tract after meals and the increased flow to muscles during exercise are examples of changes in distribution of blood flow controlled by variable arteriolar resistance. Peripheral arteriolar resistance, rather than having a cardiac output control function, has its physiological significance by its determination of distribution of blood flow.

 

Abnormal Circulation

Now that we have amalgamated the ten unique characteristics into an integrated concept of normal cardiovascular function, we proceed to examine the implications of those ten unique characteristics during pathological states.

 

HEART FAILURE (PUMP ENERGY FAILURE)
CONTRASTED WITH PUMP ENERGY EXCESS

Heart failure is the only situation where the non-sucking heart determines its output. In this state, the heart is pumping at its maximum output and therefore, in effect, by limiting the output, it determines the output.

Normally, the heart rate and pump energy are in excess of that needed to eject enough blood at systole, so that the ventricles are empty enough at diastole to allow unobstructed passive filling. Normally ventricles are not maximally filled so there is reserve compliance which allows cardiac output to be determined by the extra-cardiac factors, rather than being limited by the heart. If the heart fails, it pumps less than the normally controlling extra-cardiac factors would dictate. During pump energy failure the heart's output is being limited by whatever amount it is able to pump. The heart in such a situation of failure, by limiting output, becomes the sole regulator of cardiac output.

Definition: Heart failure exists whenever heart function is inadequate to produce the output which would be allowed by the mean cardiovascular pressure and inlet impedance. Heart failure is present when cardiac function is limiting and, therefore, determining cardiac output. Normally, extra-cardiac factors being constant, if the heart contracts more strongly, efficiently, or rapidly, no increase in output occurs. Normally, the heart is exerting excess energy over that used to produce cardiac output and thus a reserve is present, so that increases in circulation rate are not impeded by the heart. Normally, at a given circulation rate, an increase in heart rate or increase in strength of cardiac contraction does not increase the output. However, during failure where the heart is pumping out the maximal circulation rate that it can produce at the moment, increases in the rate and strength of contraction do increase cardiac output, up until venous pressure is reduced to normal, at which time the heart is no longer in failure.

When heart failure occurs, blood volume equilibrium shifts to the system behind the failing ventricle with an increase in the venous pressure at the inlet of the failing ventricle. The other ventricle, in the presence of the slowed circulation rate, is able to maintain a low inlet pressure, with the resulting shift in blood volume equilibrium toward the circuit behind the failing ventricle. With left ventricular failure, for instance, the circulation slows to the rate limited by the left ventricle and a new equilibrium occurs with a shift of blood volume from the systemic circuit to the pulmonary vascular bed. Pulmonary edema occurs if this shift in equilibrium is severe.

The high end-diastolic ventricular pressure and high end systolic volume, from a low ejection fraction in heart failure, would also be factors limiting flow, if the heart weren't already determining a lower output.

A secondary change occurs when the left ventricle fails. The lowered cardiac output decreases renal blood flow, thereby causing retention of fluid and elevation of the mean-vascular pressure. Fluid equilibrium, which exists between the cardiovascular system and extra-vascular spaces, then results in edema, liver engorgement, and the other fluid abnormalities of congestive heart failure. Heart failure can intensify in a descending spiral if the blood volume progressively expands. The over-stretching of the heart puts it at a poor mechanical advantage, thus further increasing its failure (the descending limb of Starling's curve).

An extreme example of right heart failure is seen after a "Glen" right atrial-pulmonary artery anastomosis procedure for palliation of tricuspid atresia. Here, the person has virtually complete failure of a non-functioning right heart, with a surgical bypass of the right ventricle so that, in effect, the circulation is by the left ventricle alone. Circulation is slowed by the increase in impediment that normally would be overcome by the right ventricle. As a consequence of the slowed circulation, the kidneys receive slow flow, causing retention of fluid, resulting in a permanent elevation in the homeostatic level of mean cardiovascular pressure, resulting in some compensatory increase in cardiac output at the new equilibrium.

Causes of Heart Failure:

Cardiac causes (low myocardial power):

  • Myocardiopathies � viral, arteriosclerotic, toxic, rheumatic
  • Incompetent leaky valves, e.g., mitral or aortic
  • Abnormally slow heart rate, e.g., complete heart block

Peripheral vascular causes:

  • Abnormally high mean cardiovascular pressure causing potentially higher circulation rate than a normal heart is able to deliver, e.g., renal shutdown resulting in high output failure
  • Abnormally low impedance to the heart, e.g., arteriovenous fistula

Signs and Characteristics of Heart Failure:

  1. High venous pressure is the major pathognomonic finding in cardiac failure. If the heart were not in failure, the pressure would be normally low.

     

  2. A new equilibrium occurs between the two vascular systems with a shift in volume to the field behind the failing ventricle. I.e., left ventricular failure shifts the equilibrium with a greater volume in the pulmonary circuit, resulting in x-ray evidence of pulmonary vascular engorgement and high pulmonary venous pressure. With right heart failure, the neck vein pressure is seen to be elevated. However, the systemic system initially is large enough to accommodate the shift in blood volume from the lungs without significant venous engorgement until secondary changes (as seen below) occur.

     

  3. With either left or right heart failure, increased fluid accumulation in the body progresses in a descending spiral caused by slowed cardiac output —› less than optimum renal blood flow —› low renal output —› retained fluid in the vascular system —› increased mean cardiovascular pressure —› edema —› liver engorgement and ascites.
NORMAL HEART FAILURE
Heart energy excess Heart energy deficit
Low ventricular inlet pressure High inlet pressure
Incompletely filled ventricles Completely filled ventricles
Increase in rate —› no increase in cardiac output Increase in rate —› increase in cardiac output
Increase contraction —› no increase in cardiac output Increase contraction —› increase in cardiac output
Cardiac output = f mcvp/inlet impedance Heart determines cardiac output
Body water equilibrium Progressive water retention
Normal cardiac output Low cardiac output

Pathologic Conditions Resulting in Heart Failure:

  • Complete heart block with slow pulse
  • Extremely rapid pulse
  • Intermittent, pulsatile venous flow
  • Low ejection fraction
  • Myocardiopathy with weak myocardial contraction
  • Myocardial infarction to the point of low ejection volume
  • Valvular heart disease with stenosis or insufficiency
  • Myocardial failure
  • Pericardial tamponade
  • Constrictive pericarditis

Goals in the Treatment of Heart Failure:

  1. Make the heart contract stronger and faster.
  2. Lower the resistance to ejection of blood from the ventricles.
  3. Lower the mean cardiovascular pressure by salt and water restriction and diuretics
  4. Restore the atrial effect if it is compromised by atrial fibrillation or nodal rhythym.
If failure is due to myocardial contractility, lessening the failure state by lowering mean cardiovascular pressure to the point of producing normal venous pressure, without increasing myocardial energy, may create hypovolemic shock rather than restoring normal circulation.

 

SHOCK

Shock is a low cardiac output state that results from low mean cardiovascular pressure. The low mean cardiovascular pressure can result from altering of one or the other of the two determinants of that pressure: (1) Increasing vascular compliance, and (2) Loss of blood volume.

  1. Low output with resulting hypotension from sudden loss of vascular tone is frequently seen after inducing spinal anesthetics. A dramatic example was seen following total central nervous system anesthesia, induced when marcaine was inadvertently injected into the spinal canal during an attempt to do an intercostal nerve block. In this case, sudden total relaxation of the entire vascular system resulted in severe hypotension. Loss of vascular tone can be reversed by the use of vasopressors and the addition of fluid.

     

  2. Shock from loss of blood volume can be temporarily compensated for by decreasing vascular compliance by vasopressor drugs, but ultimately needs to be corrected by blood volume restoration.

A combination of the two mechanisms of shock is seen in anaphylaxis, where there is both relaxation of the vascular system plus a shift in blood volume to the extra-vascular space, caused by increased capillary porosity.

Whatever the cause, the resulting low mean cardiovascular pressure in shock can be restored to normal by rapid infusion of fluid and electrolytes, aided by the use of vasopressor drugs.

During both of the shock mechanisms, the heart is already spending excess energy, so increasing the heart rate or strength of contraction will not increase the cardiac output. The obvious increase in output that follows the use of epinephrine, or other vasopressors, in such states is not caused by the inotropic effect on the heart, but by their effect of increasing the mean cardiovascular pressure by decreasing vascular compliance. Thus another temporary therapeutic measure, which increases mean cardiovascular pressure by making the system less compliant, is the surrounding of the vascular system by a "G-suit" such as those used by emergency medical personnel. It matters little how fast or how strongly the heart contracts: If there is no pressure in the cardiovascular system, there can be no circulation.

 

ARTERIAL HYPERTENSION
(LOW AND HIGH OUTPUT TYPES)

Arterial hypertension results from two radically different mechanisms. One is caused by increased arteriolar resistance in a system with normal mean cardiovascular pressure. The other results, in the presence of normal arteriolar resistance, from elevated mean cardiovascular pressure.

  1. In arterial hypertension from increased arteriolar resistance, there is no slowing of the circulation rate from the resistance. The heart, which normally expends excess energy, just forces whatever blood that comes to it right past the resistance point. The only change in circulation is the elevated arterial pressure. Such arterial abnormal resistance may have its origin from transient neuro-humeral stimuli. However, with longstanding arterial hypertension, changes eventually occur, resulting in more permanent structural resistance increase. If, eventually, the vascular changes effect blood flow to the kidneys, the second type of hypertension may develop.

     

  2. High mean cardiovascular pressure hypertension: The most graphic example of this state is seen during renal shutdown, from whatever cause. Here, if fluid and electrolyte intake continues, and there is reduced fluid leaving the body, the vascular volume progressively rises, with a corresponding increase in mean cardiovascular pressure. If the heart is strong enough that it continues to exert excess energy over that necessary to maintain a low ventricular inlet pressure, cardiac output increases. With the increased flow, even in the presence of unaltered arteriolar state, the resulting arterial pressure is elevated. If the heart is normal, but the pressure becomes so high that venous pressure rises, high output heart failure occurs. Severe high output failure has been seen where the systemic arterial pressure was 350/200, the systemic venous pressure was elevated with the neck veins at 30 cm., the pulmonary artery pressure was doubled, and the pulmonary capillary pressure was so high as to cause pulmonary edema with frothing of blood from the mouth and nose. This is in marked contrast to the distribution hypertension caused by arteriolar resistance increase.

The spectrum of hypertension states varies greatly between these two extremes and combinations of the two. Determination of the mean cardiovascular pressure helps delineate the significance of each in diagnosis. (See Appendix.)

Therapy for the two types of hypertension should be aimed at the etiology. For high mean cardiovascular hypertension, improving renal blood flow by removing renal arterial obstruction, diuretics, low salt diet, and water restriction to lower mean cardiovascular pressure or even renal transplant may be indicated. For hypertension caused by high arteriolar resistance, arteriolar vaso-relaxing drugs and changing the patient's response to stress may be effective.

 

ARTERIOVENOUS FISTULA

An arteriovenous fistula, with blood flow going directly from an artery to a vein without going through arterioles, capillaries, and venules, bypasses most of the resistance that has upstream compliance. Therefore, depending on size, a fistula can cause tremendous increase in cardiac output. If a shunt is great enough, high output failure can result. If a shunt is in an extremity, where it can be occluded by manual pressure, alternating decrease and increase of cardiac output can be observed, as the shunt is intermittently occluded and opened. This maneuver is a good demonstration of the peripheral vascular control of circulation rate.

Arteriovenous fistula shunts occur in a number of disease states. For example, blood flow is shunted from arteries to veins in thyrotoxic goiters. Rather than there being one large arteriovenous channel, there are many small shunting vessels. The blood may be flowing so rapidly through the many small shunts that a loud, continuous murmur may be heard over the thyroid gland. This arteriovenous shunting of blood flow accounts, in part, for the high cardiac output associated with toxic goiter.

 

CHANGES IN SIZE OF THE
VASCULAR SYSTEM

Because of the peripheral control of cardiac output, changing the size of the vascular bed, permanently (such as by a leg amputation) or by short term decrease (such as by cross-clamping the aorta and vena cava during surgery), causes an instantaneous decrease in cardiac output, mechanically, without any involvement or need of neurologic or humeral reflexes.

 

COMPLETE HEART BLOCK AND
FIXED RATE PACEMAKERS

Patients with heart block, with a very slow pulse rate, have maximal ventricular filling before the completion of diastole, thereby restricting venous flow. This restriction in venous flow causes high venous pressure and low cardiac output. As the rate is incrementally increased by the use of a pacemaker, pressure gradually falls, until the ventricles are no longer maximally filled at diastole. Further increase in pacemaker rate above that level causes no more increase in output and no further drop in the venous pressure, as ventricular capacity is no longer restrictive to flow. A rate of 60 produces this plateau in most adults at rest and in a recumbent position. If the patient is active, the pacemaker rate may need to go to 80 or even 100 before further increases in pacemaker rate cause no corresponding increase in cardiac output. When the rate is fixed at 80 and the patient is resting, the output goes down normally with a waste of energy expended by the heart, the excess not being used for circulation. The rate of 80 allows a great variation in activity without exceeding a need for pulse increase to prevent failure. If the pacemaker is set at 120, such a rate would probably accommodate cardiac output for any violent activity. However, at rest, even though the output goes down to normal, the 120 pulse tachycardia is uncomfortable, the excess myocardial energy waste is great, and myocardial oxygen availability may be taxed. Therefore, fixed pacemaker rates are usually set at about 76 to 80 as a compromise between excess myocardial energy spent for circulation at rest and needed cardiac energy for strenuous activity.

From pacemaker experience we conclude that: (1) The heart, when not in failure, is always expending excess energy over that necessary to produce cardiac output. (2) The ability to lower pulse rate below maximum during periods of low cardiac output has implications for energy conservation rather than cardiac output.

 

LEFT VENTRICULAR BOOSTER PUMPING

Myocardial energy failure is a frequent finding during acute myocardial infarction and for short periods of time after coming off of bypass after open heart surgery. In both of these situations, left ventricular booster pumping has proved useful during a recovery period.

A competent, passive filling, pulsatile outflow, continuous inflow assist pump, placed between the left atrium and aorta will automatically lower the left atrial pressure and restore circulation rate to normal. A passive filling booster pump, which expends excess energy, put in parallel with the failing left ventricle, will make the combined output of the two normal. The combined output of two non-sucking pumps in parallel will act as one pump which produces excess to that needed for normal cardiac output. Their combined output will be controlled by the extra-cardiac factors that normally control circulation rate. Whatever amount the ventricle is unable to pump will run automatically into the booster ventricle and be ejected into the aorta. When the heart has recovered, the left atrial and arterial blood pressure will remain the same whether the pump is on or off. One method of weaning the patient off of the booster pumping is to raise the pump a few centimeters above the left atrium. Blood will then run preferentially into the heart's ventricle and be pumped out by it, if the heart is no longer in power failure. If the ventricle is still in failure, the venous pressure will rise only those few centimeters. Then whatever the heart does not pump the booster pump will.

From cardiac booster pumping, we conclude that the non-failing heart is expending excess energy to that needed at any moment to produce its output. Expenditure of energy by a parallel passive filling booster pump cannot increase the circulation rate above that dictated by extra-cardiac determinants.

 

SUMMARY: The cardiovascular system is a closed elastic circle, containing two passive filling pumps in series with two vascular beds, systemic and pulmonary. Normally, the circulation rate made by the two pumps is controlled by mean cardiovascular pressure and inlet impedance. It is only during heart failure, when heart function is limiting the cardiac output, that the heart is regulating circulation rate.
 

Open-Heart Surgery with
Passive Filling Pumps

The advent of open-heart surgery has provided a wealth of data and observations regarding cardiovascular physiology. Furthermore, our expanding ability to correct cardiovascular defects has been a stimulus to better understand how the system works. Delineation of the essential mechanical features of the heart has allowed construction of heart replacement pumps that: (1) automatically provide normal circulation during heart bypass, with the flow rate remaining under control of normal physiological mechanisms; (2) allow research, in animals, by isolating the peripheral vascular versus cardiac effect on circulation rate from responses to various drug, humeral, and neural stimuli; and (3) provide a mechanical model having the ten unique characteristics of the cardiovascular system, for teaching and the study of circulatory phonomena.

 

THE MECHANICAL HEART
REPLACEMENT PUMPS

The pumps have the four characteristics in common with the heart that allow them automatically, without pump regulation, to reproduce circulation under control of normal physiologic mechanisms: (1) The pumps fill passively and don't suck at their inlets. (2) The output of the pumps is pulsatile. (3) The pumps have atria that allow uninterrupted inflow to the intermittent outflow pumps. (4) The capacity of the pump ventricles is greater than any anticipated diastolic filling.

Many designs of pumps with the above characteristics could be made. One of the six variations that the author has used is shown in Figure 4. The pump consists of a flat atrial-ventricular silicone rubber tube with a reinforcing flat cotton cover (Fig. 5). The flat configuration, which prevents rebound to a round cross-section after being compressed, is responsible for the passive filling characteristic. The tube is mechanically compressed, sequentially, by four plates activated by cams and lifters (Fig. 6). Two narrow plates act as inlet and outlet valves to the ventricular portion of the tube. Two wide plates act as atrial and ventricular impellers (Fig. 4). The atrial impeller compression and sequence are critical in preventing interruption of venous inflow when the ventricles are being emptied. Therefore, atrial compression is incomplete, leaving the atria with a 3/16" channel at maximum compression. The slope of the atrial cams is gentle, to prevent any sharp rise in pump atrial pressure that might interrupt venous flow. The cam's timing is such that the "compression let go" occurs just before the inlet valve closes.

 

CLINICAL SURGERY USING THE
HEART-SIMULATING PUMP

Cardiac surgery using both right and left heart bypass pumps has one significant advantage over conventional cardiopulmonary bypass: No oxygenator is used, as the lungs are left in the perfusion circuit and are provided with normal circulation. With normal pulmonary blood flow during the operation, post-operative pulmonary insufficiency is less of a problem than when the lungs are bypassed using an oxygenator.

By connecting the unique pumps in parallel to the heart's ventricles and with the pumps at heart level, cardiac function can be temporarily interrupted during surgical procedures without any interruption in normal circulation. When both the heart and bypass pumps are functioning simultaneously, the combined output remains the same as when the heart alone is pumping. Because both fill passively, the combined output remains regulated by the mean cardiovascular pressure and inlet impedance. Any amount that goes to the pumps doesn't go to the heart, and vice versa. If the heart function is then interrupted, by induced fibrillation or diversion of all the venous flow from one or both ventricles, the flow automatically goes to the pumps and circulation is uninterrupted and continues at the same rate. Subsequently, when the heart is defibrillated, or when the flow to the heart is no longer interrupted, the heart output takes over part of the circulation, with the total circulation rate still remaining the same. Weaning off of bypass is done simply by sequentially elevating the pumps a few centimeters at a time, thereby diverting more and more of the venous flow to the heart. Myocardial competence to take over the entire circulation becomes evident if the circulation rate and blood pressure remain unchanged after the incremental elevations are made. By appropriate elevation of the pumps above heart level, booster pumping can be done to maintain any desired atrial pressure in the heart during a recovery period.

The advent of open-heart surgery has provided a wealth of data and observations regarding cardiovascular physiology. Furthermore, our expanding ability to correct cardiovascular defects has been a stimulus to better understand how the system works. Delineation of the essential mechanical features of the heart has allowed construction of heart replacement pumps that: (1) automatically provide normal circulation during heart bypass, with the flow rate remaining under control of normal physiological mechanisms; (2) allow research, in animals, by isolating the peripheral vascular versus cardiac effect on circulation rate from responses to various drug, humeral, and neural stimuli; and (3) provide a mechanical model having the ten unique characteristics of the cardiovascular system, for teaching and the study of circulatory phonomena.

 

THE MECHANICAL HEART
REPLACEMENT PUMPS

The pumps have the four characteristics in common with the heart that allow them automatically, without pump regulation, to reproduce circulation under control of normal physiologic mechanisms: (1) The pumps fill passively and don't suck at their inlets. (2) The output of the pumps is pulsatile. (3) The pumps have atria that allow uninterrupted inflow to the intermittent outflow pumps. (4) The capacity of the pump ventricles is greater than any anticipated diastolic filling.

Many designs of pumps with the above characteristics could be made. One of the six variations that the author has used is shown in Figure 4. The pump consists of a flat atrial-ventricular silicone rubber tube with a reinforcing flat cotton cover (Fig. 5). The flat configuration, which prevents rebound to a round cross-section after being compressed, is responsible for the passive filling characteristic. The tube is mechanically compressed, sequentially, by four plates activated by cams and lifters (Fig. 6). Two narrow plates act as inlet and outlet valves to the ventricular portion of the tube. Two wide plates act as atrial and ventricular impellers (Fig. 4). The atrial impeller compression and sequence are critical in preventing interruption of venous inflow when the ventricles are being emptied. Therefore, atrial compression is incomplete, leaving the atria with a 3/16" channel at maximum compression. The slope of the atrial cams is gentle, to prevent any sharp rise in pump atrial pressure that might interrupt venous flow. The cam's timing is such that the "compression let go" occurs just before the inlet valve closes.

 

SPECIFIC APPLICATIONS OF THE
MECHANICAL PUMPS

Coronary Bypass Surgery:

Cardiac bypass during coronary surgery is used to provide a non-moving target with decompressed ventricles and the security that adequate circulation is being maintained during manipulation of the heart. Coronary artery surgery is not an open heart procedure, as the coronary arteries are on the surface of the heart. Therefore, closed bypass is applicable. Four cannulae are used. One pump is connected from the left atrium to the aorta, and the other from the right atrium to the pulmonary artery (Fig. 7).

It is important to start the left-sided bypass before the right. If the right one is started and the heart happens to fibrillate before the left bypass is started, blood will be pumped into the pulmonary circuit while no blood is leaving it. The result would be acute pulmonary engorgement ("liver lungs") with a fatal outcome. Likewise, the left-sided bypass should be discontinued last.

The heart is electrically fibrillated while the coronary-graft anastomoses are made. Then the heart can be defibrillated before the aorta-graft anastomoses are made. If, during the coronary-graft anastomoses, greater decompression of the ventricles is desired, the atrial cannulae can be advanced into the ventricles to drain them as well as the atria. Figure 8 shows arterial blood pressures staying relatively constant before and during a coronary bypass procedure, using right and left heart bypass.

During cardiac bypass with the two pumps, with the lungs functioning and without the use of an oxygenator, the anesthesiologist maintains ventilation and support of circulation just as he would in other non-cardiac procedures. Circulation rate reacts to blood loss, fluid infusion, and vasoactive drugs, just as when the heart is functioning.

Pulmonary Stenosis:

While it might appear that pulmonary stenosis could be repaired by using a right-sided bypass alone, one laboratory experience has demonstrated this to be too hazardous for clinical application. During the course of a right heart bypass, with the left ventricle functioning, the heart unexpectedly fibrillated. With the left ventricle suddenly not pumping, and before the right heart bypass pump could be turned off and the heart defibrillated, the lungs became irreparably overloaded with blood. The resulting "liver lungs" were not reversible, causing the death of the animal. Therefore, for safety, total heart bypass is always used, even when a right heart bypass alone could allow adequate access for correction of the lesion.

Use of a Passive Filling Pump with a Bubble Oxygenator:

During open-heart surgery, two pump bypass is precluded because of the difficulty in bypassing the left atrium, with its many pulmonary veins. Therefore, cardiopulmonary bypass with an oxygenator is necessary whenever the cardiac chambers need to be opened.

The passive filling, pulsatile output, continuous inflow pump has advantages over other types of pumps when used in a pump-oxygenator circuit. It can automatically produce normal circulation rate with a normal pulse wave, without any control adjustments, and without the hazard of oxygenator blood level fluctuations or air emboli.

If the inlet of the pump is placed at the priming level in the oxygenator (Fig. 9), because the pump is non-sucking, the blood will never fall below that level. Any blood that runs into the oxygenator that would tend to raise the level is automatically pumped back into the patient. The oxygenator is positioned at such a level that normal venous pressure is maintained by gravity drainage during bypass.

With this pump-oxygenator setup, circulation rate is determined by the patient's mean cardiovascular pressure and inlet impedance, just as it does in the intact body. During the bypass, circulation rate is modified by using vasoactive drugs, blood and fluid replacement to change the mean cardiovascular pressure, not by pump alteration.

This technique — beginning partial bypass in parallel with the heart — results in normal circulation rate, as both the heart and the extra-corporeal system fill passively. When complete bypass is produced, by occluding the vena cava around the caval catheters, the circulation rate remains the same. Terminating bypass is very simple. The occluding tapes around the vena caval tubes are released and the venous drainage tubing is occluded in increments. As more and more blood is diverted to the heart, the venous pressure will indicate whether or not further decrease in extra-corporeal pumping will be tolerated. In this way, the pump-oxygenator can be used as an automatic booster device during the myocardial reovery period, following the cardiac repair procedure. Figure 10 shows the blood pressure before, during, and after bypass, including the parallel bypass at the end of the procedure, in a patient with wide-open aortic insufficiency. The wide, abnormal pulse wave of aortic insufficiency is followed with a normal pulse wave while on bypass and after valve replacement. Figure 11 shows typical pulse waves during bypass and Figure 12 shows superimposed waves during circulation produced by parallel pumping of the heart and extra-corporeal pump.

Figure 13 illustrates a safety feature of the passive filling pump used with an oxygenator. Bypass was just underway, during a mitral valve replacement, when the inferior vena cava cannula slipped back into the right atrium. The inferior vena cava being temporarily occluded by the circumferential tape, caused the venous drainage to the pump to drop to one liter per minute. The oxygenator did not run out of blood and no air was pumped into the patient during this low output episode. After the tube was re-advanced properly into the inferior vena cava, the blood flow to and from the pump automatically returned to normal.

The Non-Sucking Pump Used with a Membrane Oxygenator:

The non-sucking heart replacement pump is ideal for use with a membrane oxygenator. It automatically allows a constant blood volume in the system with this closed-volume oxygenator. There is no need for a reservoir for volume monitoring of changing blood levels, which would occur and require alteration of pump rate with other types of pumps. In effect, the passive filling, continuous inflow pump used with a membrane oxygenator provides a closed automatic bypass system.*

 

*NOTE: In all clinical applications of passive filling pumps, it is important to use large caliber venous drainage tubing so as not to add to the vascular system's normal inlet impedance to the pumps.

Pulsatile Blood Flow During Bypass Surgery:

Cardiac bypass during heart surgery has given an opportunity to observe the benefit of pulsatile blood flow, by comparing it with techniques using non-pulsatile flow. Pulsatile flow ensures diffuse normal distribution of blood flow to all the organs and tissue of the body, while non-pulsatile flow results in reduced flow to certain vascular beds and excessive flow to others. With non-pulsatile flow, the brain and kidneys receive reduced blood supply. Also, incomplete metabolism from islands of under-perfused tissue results in acidosis, not found when pulsatile flow is provided. When hypothermia is used, the need of pulsatile flow to insure diffuse distribution of circulation is increased. While the adverse effects of non-pulsatile flow can be offset by the use of vaso-relaxers, blood dilution, and other methods, the value of pulsatile flow in normal function is illustrated by comparing findings from the two types of pumping systems.

Ventricular Compliance — an Inlet Impedance Factor:

Open-heart surgery on a patient with severe hypertrophy of the left ventricle, from longstanding aortic stenosis, illustrated how ventricular compliance can be a factor in the inlet impedance determination of cardiac output. This patient's left ventricle measured three-fourths of an inch in thickness. After replacement of the aortic valve, the bypass was discontinued. The heart contracted very strongly, with a normal pulmonary venous pressure of 12 cm. water. However, there was practically no cardiac output and the blood pressure remained at only 55/25. The blood volume was, therefore, increased by increments while watching the pulmonary venous pressure go to 16, 18, 20, 25, and 30 cm., with no effect on output or blood pressure. The heart continued to beat very forcibly. Suddenly, at a pulmonary venous pressure of 35 cm., almost at the point of causing pulmonary edema, the heart distended during diastole, cardiac output went beyond normal with a resulting blood pressure of 160/90. The non-compliance of this thick ventricle was a significant impediment to passive filling and cardiac output.

Hydraulic Model of the Cardiovascular System

A model, incorporating many of the unique characteristics of the cardiovascular system, allows the opportunity to observe the significance of each of those features. The characteristics reproduced in the model are: (1) the hydraulic system is a circle; (2) it is an elastic system, (3) it is filled with fluid producing a mean pressure; (4) there are two pumps in series between two vascular beds; (5) the pumps fill passively; (6) pump output is intermittent; (7) the pumps have atria that allow continuous, unimpeded flow to the pumps; (8) the pumps are capable of pumping out a greater volume than the system will produce; and (9) there are resistance points near the pumps' inlets as well as near their outlets.

Studying the model has an advantage over studying in vivo preparations in that the effect of altering one variable at a time can be directly observed. Furthermore, the model eliminates the possibility that some simultaneous, hidden, undetected change might have occurred that would invalidate cause and effect conclusions (Fig. 37). The model has proven to be a very effective study and teaching tool.

The Model:

Figure 38 shows the layout of the model pictured in Figure 39.

The Pumps:

The model pumps share the three unique characteristics of the heart: They are non-sucking, and therefore fill passively at their inlets, they have a pulsatile outflow, and they have atria which allow uninterrupted inflow to the intermittent, pulsatile outflow ventricles.

Figure 40 shows one of the two assembled pumps. Figure 41 shows the unassembled parts of a pump.

The atrial and ventricular pump chambers consist of a rigid side (Fig. 41 at A) and an opposing pliable silicone rubber side (Fig. 41 at B). The ventricular half of the rigid side has both an inlet and an outlet port containing inlet and outlet valves, respectively. The atrial half of the rigid side has a single port with no valve. There is a "Y" connection between the atrial port, inlet port to the ventricle and the venous line. This "Y" inlet allows venous fluid to run both into the atrium, when the inlet ventricular valve is closed during systole, and into the ventricular chamber during both ventricular diastole and atrial systole. This arrangement reproduces the "atrial effect" by allowing continuous uninterrupted flow from the veins to the intermittent outflow ventricles, thus preventing the need of overcoming inertia by starting stopped flow after each pump beat.

The impeller part of the pump also consists of a rigid side (Fig. 41 at E) and an opposing pliable rubber side (Fig. 41 at D). Each half of the rigid part has a port through which air can be forced into or sucked out of the impeller chambers. When the impeller chambers are inflated, their rubber side exerts pressure on the silicone rubber side of the ventricle or atrial chambers, thus emptying the ventricle or atrium.

A spacer (Fig. 41 at C), placed between the opposing impeller and pumping chambers, has many side holes which allow air to pass freely to and fro between the ambient air and the space between the impeller and pump ventricles. This free communication prevents any negative pressure, caused by suction used to deflate the impeller spaces, from being transmitted to the pump ventricles. Thus, the ventricles fill passively from venous pressure.

The power supply to the pumps is from compressed air and vacuum sources, delivered alternately to the impeller spaces timed in the desired sequence. An electronic regulator triggers solenoid valves to control pulse rate and the ratio of systole to diastole. The pulse rate is variable from 1 to 160 strokes per minute at an impeller pressure of one to thirty p.s.i. Reducer valves regulate peak pressure and onset to peak time. Impeller deflating suction is strong enough that passive ventricular filling is not impeded.

Simulated Vascular Network:

The major vessels are silicone tubing, while the capillary beds are made of more compliant Penrose drain material. Variable resistance points are at various sites in both arterial and venous lines. Pressure transducers connect to a direct writing recorder and pressure tubes connect to an electromagnetic flow meter. A transfusion reservoir provides opportunity to change the fluid volume within the system. Different weights placed on a plate resting on the "capillary beds" allow changes in the system compliance to be made. The system was filled with water. Hundreds of pieces of #0 silk suture material, 2 millimeters in length (found to be isobaric with water), were suspended in the water to allow visualization of the circulation.

 

MODEL FINDINGS CORRELATE WITH OBSERVATIONS IN HUMAN PHYSIOLOGY:

 

EXPERIMENT #1
The relationship of mean cardiovascular pressure to volume in the system

With the pumps turned off, the compliance remaining unchanged, and starting with the system full of water, at a volume of 2000 cc. at a pressure of zero, pressure was recorded as fluid was added by way of the transfusion reservoir. Figure 43 shows the non-linear relationship between the mean system pressure and the volume.

 

EXPERIMENT #2
The relationship of circulation rate to mean system pressure

Starting with a mean system pressure of zero, and with arbitrary settings of the resistance points, the pumps were started. Appropriate amounts of fluid were added to increase the mean system pressure by increments of 2 cm. water pressure (Fig. 44)

Findings:

 

  1. Figure 44 shows a direct correlation between pump output and mean system pressure over the range of 0 to 20 cm. of water pressure. The greater the mean system pressure, the greater the pumps' output.

     

  2. At no time, over the range of pressure studied, did the ventricles fill to capacity at diastole. There was always a "pump capacity excess."

     

  3. At no time was there any interruption of venous flow by the intermittent closing of the inlet valves of the ventricles at systole, as the atria had been emptied during ventricular diastole.

 

EXPERIMENT #3
Pump rate correlation with pump output

The pump rate in the model was varied from 0 to 170 beats per minute with other variables remaining constant (Fig. 45). The mean system pressure was kept at 12 cm. water pressure, with the impeller pressure at 14 p.s.i. and the resistance kept constant.

Findings:

 

  1. At pump rates between 0 and 20/min.:
    • The pump output correlated in a linear way with the pump rate (Fig. 45, from A to halfway to B).
    • The ventricles filled to capacity at each ventricular diastole.
    • The atria filled to capacity during each ventricular systole.

     

  2. At pump rates between 20 and 40/min.:
    • The pump rate correlated in a non-linear way with the pump output (Fig. 45, from halfway in between A and B, to B).
    • The ventricles filled slightly less than capacity at ventricular diastole.
    • The atria were filled to capacity at the end of ventricular systole.
    • Venous flow was slowed but not completely interrupted before the end of ventricular systole.

     

  3. Increasing the pump rate from 50 to 110:
    • Caused no increase in pump output (Fig. 45, B to C).
    • Was associated with progressive decrease in ventricular end-diastolic volume.
    • Caused progressive decrease in atrial diastolic volume.
    • Was unaccompanied by any interruption in venous flow.
    • The pump rate and stroke volume were reciprocals of one another.

     

  4. Increasing the pump rate from 110 to 170 beats/min. (Fig. 45, C to D) was accompanied by a progressive increase in residual air pressure in the ventricular impeller after ventricular systole which:
    • Progressively decreased pump output.
    • Impeded ventricular filling at ventricular diastole.
    • Caused venous flow interruption at ventricular systole.

 

EXPERIMENT #4
Ventricular impeller-force correlation with pump output

The ventricular impeller pressure was varied from 0 to 20 p.s.i., in increments of 2 p.s.i., while keeping other variables constant (Fig. 46). The pump rate was 80, the mean system pressure was 12 cm. water pressure and all resistances remained unchanged.

Findings:

 

  1. At impeller pressures between 0 and 10 p.s.i. there was:
    • Linear correlation between impeller pressure and pump output (Fig. 46, from A to B).
    • The ventricles were never completely emptied at ventricular systole.
    • The ventricles were completely filled at ventricular diastole.
    • Venous flow to the pumps was interrupted at each ventricular systole.

     

  2. At impeller pressures between 10 and 18 p.s.i. there was:
    • No increase in the output as the pressure increased (Fig. 46, from B to C).
    • Maximal ventricular emptying at systole.
    • Sub-maximal filling of the ventricles at diastole
    • No venous flow interruption at any time.

     

  3. As the impeller pressure was progressively increased above 18 p.s.i. (Fig. 46, from C to D) there was progressively incomplete evacuation of compressed air in the ventricle during diastole which:
    • Progressively decreased pump output.
    • Impeded ventricular filling at diastole.
    • Caused venous flow interruption at ventricular systole.

 

EXPERIMENT #5
Relationship of resistance to pump output

Resistance at a variety of sites was changed in the model's vascular network while maintaining other factors constant, with the mean system pressure at 12 cm. water pressure, pump rate at 80 beats per minute, and an impeller pressure of 16 p.s.i. There was a marked difference in response to a given resistance depending on where the resistance was placed in the model. Responses fell into two categories which became more obvious the closer the resistance was to either the inlet or outlet of a pump. Therefore, resistance was studied in two situations: near a pump inlet, where there was a large compliant bed upstream; and near a pump outlet, with no compliant bed upstream.

Relationship of Resistance Near a Pump Outlet to Pump Output: (Fig. 47)

A variable resistance clamp was placed 20 cm. from the left heart homologue pump outlet (Fig. 38, #9), across which a pressure gradient could be monitored by transducers (Fig. 38, #8).

 

Findings:
  1. As the resistance is progressively increased (Fig. 38, #9), the superimposed arterial pressure tracings, which demonstrated no gradient initially (Fig. 47 at A), separated as resistance was added (Fig. 47, from A to C).
  2. With increase in outflow resistance up to 100 mm. Hg. gradient, there was no drop in pump output (Fig. 47 at C).
  3. Further increase in gradient from 100 to 250 mm. Hg did cause a corresponding drop in flow (Fig. 47, from C to D).
  4. In the range where there was no drop in flow (Fig. 47, A to C), the left ventricle did not fill maximally, but did empty maximally, and there was no interruption of venous flow at any time during the pumping cycle.
  5. However, when the gradient became great enough that the flow slowed (Fig. 47, C to D), the left ventricle progressively emptied less completely, and the residual volume resulted in complete left ventricular filling at diastole and interruption of venous inflow at each systole.
  6. Whenever the left ventricle was being filled completely at diastole, increasing the mean system pressure caused no concomitant increase in pump output.
  7. As slowing of flow occurred (Fig. 47, C to D), there was gradual increase in a greater than previous volume in the pulmonary circuit at the expense of the systemic circuit. The greater the slowing, the more the volume equilibrium shifted to the simulated pulmonary circuit.

The Effect of Resistance Near a Pump Inlet on Pump Output: (Fig. 48)

An adjustable resistance clamp was placed 20 cm. from the left pump's inlet (Fig. 38, #7), across which a pressure gradient could be monitored using the transducers (Fig. 38, #6). The clamp was progressively closed. The resulting inflow resistance magnitude is represented by the pressure gradient between the transducers. Figure 48 shows the two superimposed pressure lines (Fig. 48, A to B) in the absence of resistance. When resistance is progressively increased, starting at B, a pressure gradient becomes evident (Fig. 48, B to C).

 

Findings:
  1. There was a direct inverse correlation between pump output and inflow resistance to the pump (Fig. 48, B to C) over the whole range studied. This is in marked contrast to outflow resistance of the previous experiment (Fig. 47, B to C), where decrease in flow did not occur until there was pump failure.
  2. As inlet resistance increased there was progressively less filling of the ventricles at diastole. Maximal emptying did occur at each ventricular systole, and venous flow remained uninterrupted at all times.
  3. Progressive increase in left pump inflow resistance shifted the fluid equilibrium between the pulmonary and systemic circuits toward the pulmonary bed.
  4. The circuit with the greatest inflow resistance was the major determinant of the output of both.

 

EXPERIMENT #6
The Atrial Effect

In the model, to eliminate the atrial effect, the air pressure in the atria can be left continuously obliterating the atrial chambers, while not impinging on the connection of the venous tubes to the ventricles. By intermittently producing circulation with and without the atrial effect, its contribution to pump output was determined. At pump rates between 60 and 90 there was four times the pump output (2400 cc./min.) with the atria functioning as with no atrial effect (600 cc./min.). The output without atria progressively decreased as the rate was increased. At 130 beats per minute, flow stopped completely. It is demonstrated that atria markedly increase flow when used with pulsatile, passively filling pumps. They prevent inertia, which would otherwise occur if the inlet valves were allowed to stop venous flow at each beat.

 

EXPERIMENT #7
Pulsatile Flow

A branching hydraulic system was made with twenty, 3/16 inch diameter, transparent tubes, six to twelve inches in length, linked together with "Y" connectors in the configuration of a simulated vascular bed.

Findings:

 

  1. With pulsatile flow there was flow in all parts of the system. It fairly danced with the pulsation. There were areas where flow was in one direction during most of the cycle and reversed in the other direction for the rest of it. With pulsatile flow, the distribution remained stable for a long period of time.

     

  2. With continuous non-pulsatile flow, the distribution started out in a diffuse manner. Very shortly, some channels had higher flow than others. Yet there was some flow in all of them. However, within a short time the 2 mm. silk particles began clustering at "Y" junctions, eventually totally blocking some and partially blocking others. Before long, almost all of the flow was in just a few channels and very little was in the others. When pulsatile flow then replaced the steady flow, the aggregates of silk were broken up, the whole bed began dancing with the pulse, and flow was again established in a diffuse manner, with flow fairly equal in all channels.

Conclusion:

 

There is obviously a difference in the mechanism of aggregation of silk particles in this system and the mechanism that caused poor distribution with non-pulsatile flow in the animal in the previous chapter. However, this experiment does show that the simplest way to guarantee diffuse, equal distribution of fluid in a hydraulic system is to make flow pulsatile.

 

THE MODEL AS A TEACHING TOOL

An example of the teaching and learning benefit of the model is illustrated by an incident. One day I had scheduled a replacement of an aortic and mitral valve in a woman who had severe mitral and aortic stenosis. The student on my service came to me and said, "This woman is so critically ill that she may have difficulty surviving such a big operation. Maybe it would be the better part of valor to correct only one valve. Then, when she is better, go back and replace the other one." I asked him which valve he thought I should replace first. He responded that he would go to the model and put an inlet and outlet obstruction on the circuit. Then remove one and then the other and see what improvement would result from a single obstruction removal. When he only removed the inlet obstruction, the increased flow to the pump put it into power failure. When he removed only the outlet obstruction it decreased the workload of the pump, but the output didn't increase.

We would have sealed the woman's fate if we had corrected only one of the valves. She would have remained in low output failure if we had corrected the aortic valve alone; and she would have gone into acute left ventricular failure if we had relieved the mitral stenosis alone. Therefore, we replaced both valves and the patient made an excellent recovery.

 

The close correlation of model findings to physiological observations in man makes the model useful in understanding how the cardiovascular system works, and helps to anticipate cause and effect in human physiology.
 

The significance of ten unique characteristics of the cardiovascular system has been demonstrated by clinical observations, cardio-bypass data, animal mechanical heart replacement experiments, and a simulated cardiovascular model. All of the evidence validates that:

 

  1. The cardiovascular system is a circle containing a mean cardiovascular pressure dependent upon volume and compliance.

     

  2. The heart is a passively filling pump.

     

  3. Atria increase cardiac output by causing continuous venous flow to the intermittently contracting ventricles, not by pumping up the ventricles.

     

  4. Circulation rate is normally determined by the extra-cardiac factors: mean cardiovascular pressure and inlet impedance.

     

  5. It is only during heart failure that the heart is the regulator of cardiac output.

 

Clinical Determination of Mean-Cardiovascular Pressure

A useful approximation of the mean-cardiovascular pressure can be made without the need of stopping the heart and letting the pressure equalize in the entire cardiovascular system. The assumption is made that the same pressure would be found by isolating a representative sample of the arteries, capillaries, and veins at ventricular diastole, and letting the pressure equilibrate in that sample. By instantaneously interrupting arterial inflow to the arm and venous outflow from it, the pressure will fall in the arteries and rise in the veins until they are equal. This equalized pressure, which will occur within 30 seconds, approximates that found when circulation is stopped in the entire body.

Equipment:

A narrow pneumatic blood pressure cuff (approximately one inch in width) is used so that during inflation it will not displace any blood volume distally into the arm.

A one liter air pressure reservoir, pressurized to 300 mm. Hg, is connected to the pressure cuff with a valve interposed to allow instantaneous inflation of the cuff. The reservoir is pressurized by a regular blood pressure inflating bulb and valve, interposed with a "Y" connecter between the reservoir and the valve.

Two pressure transducers and recorders are needed, to allow pressure recording without any loss of fluid from the vascular bed, which would occur with a simple manometer system.

Technique:

     

  1. The patient should be lying perfectly horizontal.

     

  2. Both an artery and a vein are cannulated at the antecubital area of the arm and connected to the pressure transducers.

     

  3. The blood pressure cuff is applied above the biceps bulge of the arm. It must be applied loose enough that it does not cause any venous obstruction, as evidenced by the observation that it produces no elevation of venous pressure above that seen before its application.

     

  4. The cuff must be applied tight enough that its inflation, to 300 mm. Hg pressure, completely interrupts the arterial flow to the arm. Complete interruption can be assumed if arterial and venous pressure approach equalization after 30 seconds, and do not continue to rise thereafter.

     

  5. The arm should be abducted sufficiently from the side of the body, so as to obtain the lowest venous pressure reading, thereby being certain that any abduction is not mechanically interfering with venous flow in the axilla.

     

  6. The arm should be horizontal, with the anterior surface of the antecubital skin at mid- chest position.

     

  7. The thickness of the chest should be accurately measured. The pressures are small, so careful standardization is necessary in order to get meaningful data.

     

  8. The patient is instructed to leave the arm perfectly relaxed, and not to contract any muscle.

     

  9. The valve or switch in the pressure line is turned, thereby inflating the cuff, which simultaneously interrupts venous and arterial flow to and from the arm.

Pressure readings are made 30 seconds after the occlusion. As the equalization process progresses, the pressure gradient becomes so small between the arteries and veins that the pressures may never completely equalize. However, the average of the arterial and venous pressures recorded at 30 seconds can be interpreted as the mean-cardiovascular pressure.

Establishing such a standardization makes these approximate readings clinically useful. In any low arterial blood pressure situation, the mean-cardiovascular pressure indicates whether the problem is from low blood volume or myocardial failure. Mean-cardiovascular pressure is a diagnostic tool that can separate high output hypertension from the more common arteriolar resistance hypertension, thus indicating the proper therapy regimen.

 

Function of the cardiovascular system

Blood circulates through a network of vessels throughout the body to provide individual cells with oxygen and nutrients and helps dispose of metabolic wastes. The heart pumps the blood around the blood vessels. Blood is made up of about 45% solids (cells) and 55% fluids (plasma). The plasma is largely water, containing proteins, nutrients, hormones, antibodies, and dissolved waste products.

Functions of blood and circulation:

  • Circulates OXYGEN and removes Carbon Dioxide.
  • Provides cells with NUTRIENTS.
  • Removes the waste products of metabolism to the excretory organs for disposal.
  • Protects the body against disease and infection.
  • Clotting stops bleeding after injury .
  • Transports HORMONES to target cells and organs.
  • Helps regulate body temperature.

 


Blood

General types of blood cells: (each has many different sub-types)

ERYTHROCYTES (red cells) are small red disk shaped cells. They contain HAEMOGLOBIN, which combines with oxygen in the lungs and is then transported to the body's cells. The haemoglobin then returns carbon dioxide waste to the lungs. Erythrocytes are formed in the bone marrow in the knobby ends of bones.

LEUKOCYTES (white cells) help the body fight bacteria and infection. When a tissue is damaged or has an infection the number of leukocytes increases. Leukocytes are formed in the small ends of bones. Leukocytes can be classed as granular or non granular. There are three types of granular leukocytes (eosinophils, neutrophils, and basophils), and three types of non-granular (monocytes, T-cell lymphocytes, and B-cell lymphocytes). See also the lymphatic system.

THROMBOCYTES (platelets) aid the formation of blood CLOTS by releasing various protein substances. When the body is injured thrombocytes disintegrate and cause a chemical reaction with the proteins found in plasma, which eventually create a thread like substance called FIBRIN. The fibrin then "catches" other blood cells which form the clot, preventing further loss of blood and forms the basis of healing.

 


Blood vessels

ARTERIES carry oxygenated blood away from the heart. They are thick hollow tubes which are highly ELASTIC which allows them to DILATE (widen) and constrict (narrow) as blood is forced down them by the heart. Arteries branch and re-branch, becoming smaller until they become small ARTERIOLES which are even more elastic. Arterioles feed oxygenated blood to the capillaries. The AORTA is the largest artery in the body, taking blood from the heart, branching into other arteries that send oxygenated blood to the rest of the body.

CAPILLARIES distribute the nutrients and oxygen to the body's tissues and remove deoxygenated blood and waste. They are extremely thin, the walls are only one cell thick and connect the arterioles with the venules (very small veins).

VENULES (very small veins) merge into VEINS which carry blood back to the heart. The vein walls are similar to arteries but thinner and less elastic. Veins carry deoxygenated blood towards the lungs where oxygen is received via the pulmonary capillaries. The PULMONARY Veins then carries this oxygenated blood back to the heart.

 


The heart

The heart is a hollow muscular organ which beats over 100,000 times a day to pump blood around the body's 60,000 miles of blood vessels. The right side of the heart receives blood and sends it to the lungs to be oxygenated, while the left side receives oxygenated blood from the lungs and sends it out to the tissues of the body. The Heart has three layers; the ENDOCARDIUM (inner layer), the EPICARDIUM (middle layer), and MYOCARDIUM (outer layer). The heart is protected by the PERICARDIUM which is the protective membrane surrounding it.

The heart has FOUR CHAMBERS, in the lower heart the right and left Ventricles, and in the upper heart the right and left Atria. In a normal heart beat the atria contract while the ventricles relax, then the ventricles contract while the atria relax. There are VALVES through which blood passes between ventricle and atrium, these close in such a way that blood does not backwash during the pauses between ventricular contractions. The right and left ventricles are divided by a thick wall (the VENTRICULAR SEPTUM), babies born with "hole in the heart" have a small gap here, which is a problem since oxygenated and deoxygenated can blood mix. The walls of the left ventricle are thicker as it has to pump blood to all the tissues, compared to the right ventricle which only pumps blood as far as the lungs.

 


The spleen

This is a large flat oval organ located below the diaphragm, it's main function is to STORE BLOOD. The size of the spleen can vary, for example it may enlarge when the body is fighting infection also it's size tends to decrease with age. It is a non-vital organ and it is possible to survive after removal of the spleen.

Perinicious anaemia is a Vitamin B12 deficiency resulting in a reduction in number of erythrocytes.

Aplastic anemia is a failure of the bone marrow to produce the enough red blood cells.

Septicaemia - bacterial toxins in blood.

 


Roots, suffixes, and prefixes

component meaning example
CARDIO- heart echocardiogram = sound wave image of the heart.
CYTE- cell thrombocyte = clot forming cell.
HAEM- blood haematoma - a tumour or swelling filled with blood.
THROMB- clot, lump thrombocytopenia = deficiency of thrombocytes in the blood
ETHRO- red ehtrocyte = red blood cell
LEUKO- white leukocyte = white blood cell
SEP, SEPTIV- toxicity due to micro-organisms septicaemia
VAS- vessel / duct cerebrovascular = blood vessels of the cerebrum of the brain.
HYPER- excessive hyperglycaemia = excessive levels of glucose in blood.
HYPO- deficient / below hypoglycaemia = abnormally low glucose blood levels.
-PENIA deficiency neutropenia = low levels of neutrophilic leukocytes.
-EMIA condition of blood anaemia = abnormally low levels of red blood cells.

 


Cancer Focus

 

Overview of Haematological Malignancies
 
The most common haematological malignancy is leukaemia - cancer of the white blood cells. There are many types of leukaemia; Acute types progress rapidly, while Chronic types develop more slowly. Leukaemia is often accompanied by anaemia because the red oxygen carrying cells in the blood are crowded out by the cancerous white cells. There are a number of malignancies and disorders affecting other types of blood cells.
Internet Resources for Leukaemia

 

Acute Lymphoblastic Leukaemia (ALL)
 
Acute lymphoblastic leukaemia (also known as acute lymphocytic leukaemia or ALL) is a disease where too many immature lymphocytes (a type of white blood cell) are found in the blood and bone marrow. Symptoms can include persistent fever, weakness or tiredness, achiness in the bones or joints, or swollen lymph nodes. Adult ALL and its treatment is usually different to childhood ALL. Almost a third of adult patients have a specific chromosome translocation; "Philadelphia Positive" ALL.
Internet Resources for Acute Lymphocytic Leukaemia

 

Acute Myeloid Leukaemia (AML)
 
Acute myeloid leukemia (AML) is a disease in which too many immature granulocytes (a type of white blood cell) are found in the blood and bone marrow. There are a number of subtypes of AML including acute myeloblastic leukemia, acute promyelocytic leukemia, acute monocytic leukemia, acute myelomonocytic leukemia, erythroleukemia, and acute megakaryoblastic leukemia.
Internet Resources for Acute Myeloid Leukaemia

 

Other Types of Leukaemia
 

Chronic Lymphocytic Leukaemia
Chronic Myelogenous Leukaemia
Hairy Cell Leukaemia
Internet Resources for Leukaemia

 

Childhood Leukaemia
 
Childhood leukaemias tend to have different characteristics and treatments compared to adult leukaemias. There is a "childhood peak" of Acute Lymphoblastic Leukaemia, there is a lower proportion of Acute Myeloid Leukaemias compared to adult patients. Clinical prognostic factors include age, White Blood Cell count (WBC) at presentation, and Central Nervous System (CNS) involvement. Infants less than 1 year and adolescents over 10 years of age, WBC greater than 50,000, or CNS involvement are associated with a less favourable prognosis.
Internet Resources for Childhood Leukaemia

 

Other Haematological Malignancies
 
- Myelodysplastic Syndromes
Myelodysplastic syndromes, sometimes called "pre-leukaemia" are a group of diseases in which the bone marrow does not produce enough normal blood cells. Common symptoms are anaemia, bleeding, easy bruisability, and fatigue. These Myelodysplastic syndromes can occur in all age groups but are more common in people aged over 60. Myelodysplastic syndromes may develop spontaneously or be secondary to treatment with chemotherapy / radiotherapy. There is an association with Myelodysplastic syndromes and acute myeloid leukaemia.
- Myeloproliferative Disorders
Myeloproliferative disorders are diseases in which too many blood cells are made by the bone marrow, there are 4 main types of myeloproliferative disorders: chronic myelogenous leukaemia, polycythemia vera, agnogenic myeloid metaplasia, and essential thrombocythemia. Chronic myelogenous leukaemia is where an excess of granulocytes (immature white blood cells) are found in the blood and bone marrow. Polycythemia vera is where red blood cells become too numerous often resulting in a swelling of the spleen. Agnogenic myeloid metaplasia is a condition in which certain blood cells do not mature properly, this may result in a swelling of the spleen and anaemia. Essential thrombocythemia is a disease in which the body produces excessive numbers of platelets (cells in the blood that make it clot) which impedes the normal circulation of blood.
- Aplastic Anaemia
Anaplastic Anemia is not a cancer. AA is a rare disease in which the bone marrow is unable to produce adequate blood cells; leading to pancytopenia (deficiency of all types of blood cells). AA may occur at any age, but there is a peak in adolescence / early adulthood, and again in old age. Slightly more males than females are diagnosed with AA, also the disease is more common in the Far East. Patients successfully treated for aplastic anemia have a higher risk of developing other diseases later in life, including cancer.
- Fanconi Anaemia
Fanconi Anaemia is not a cancer, it is a rare disorder found in children that involves the blood and bone marrow. The symptoms include severe aplastic anemia, hypoplasia of the bone marrow, and patchy discoloration of the skin. Recent research has shown an association between Fanconi anaemia and leukaemia.
- Waldenstrom's Macroglobulinemia
This is a rare malignant condition, involving an excess of beta-lymphocytes (a type of cell in the immune system) which secrete immunoglobulins (a type of antibody). WM usually occurs in people over sixty, but has been detected in younger adults.
Internet Resources for Haematological Malignancies

 

French-American-British (FAB) Classification Scheme
 
Leukaemia can be classified using the French-American-British (FAB) criteria. for cell morphology:

L1 - ALL: small lymphoid cells, regular nuclei
L2 - ALL: large lymphoid cells, irregualr nuclei
L3 - ALL: large homogeneous cells with prominent nucleolus
M1 - Myeloblastic leukemia without maturation
M2 - Myeloblastic leukemia with maturation
M3 - Promyelocytic leukemia
M4 - Myelomonocytic leukemia
M5 - Monocytic leukemia
M6 - Erythroleukemia
M7 - Megakaryoblastic leukemia
M0 - AML with minimal differentiation

 

CNS Prophylaxis
 
Leukaemia can sometimes spread to the spinal cord and brain (Central Nervous System). Intrathecal chemotherapy (injection into the fluid around the spine) may be given to combat or prevent CNS relapse.

 

Blood Counts
 
Blood counts are done to test the number of each of the different kinds of cells in the blood. This may be an aid to diagnosis or done to monitor toxicity after each course of chemotherapy. The next course of chemotherapy may be delayed until white cells, neutrophils, and platelets have recovered to a safe level.

 

Cardiotoxicity
 
Cardiotoxicity (damage to the heart) is associated with certain anti cancer drugs, especially Adriamycin. As such the total dose of these drugs may be limited to reduce the risk of cardiotoxicity.

 

Echocardiagram
 
An Echocardiogramis where an image of the heart is formed when high frequency sound waves are reflected from the muscles of the heart. An echocardiogram may be done before treatment starts to establish a baseline from which to compare future tests.

 

Metastases through the cardivascular system
 
The network of blood vessels reach all parts of the body and may provide one of the routes for cancer cells to spread to secondary sites.

Related Abbreviations and Acronyms:

 

 
AA Anaplastic Anaemia
ALL Acute lymphoblastic leukaemia
AML Acute Myeloid leukaemia
ANC Absolute neutrophil count
ANLL Acute non-lymphatic leukaemia
ASH American Society for Hematology
B-ALL B-cell Acute Lymphoblastic Leukaemia
BP Blood pressure
CALGB Cancer and Leukemia Group B (USA)
cALL Common ALL
CGL Chronic Granulocytic Leukaemia
CHF Congestive heart failure
CLL Chronic lymphocytic Leukaemia
CML Chronic myeloid leukaemia
CMML chronic myelomonocytic leukemia
CPR Cardio pulmonary resuscitation
CVA Cardiovascular Accident (stroke)
CVC Central venous catheters
ECG Electrocardiogram - heart scan
FAB French American and British classification scheme for leukaemia
FBC Full Blood Count
G-CSF Granulocyte colony stimulating factor promotes production of white blood cells
GM-CSF Granulocyte and macrophage colony stimulating factor
Hb Haemoglobin
HCL Hairy Cell Leukaemia
HD Hodgkin's Disease (lymphoma)
HTLV Human T-cell leukemia-lymphoma virus
IV Intravenous - into a vein
LRF Leukaemia Research Fund (UK)
LVEF Left Ventricular Fjection Fraction - a heart function test
LVSF Left Ventricular Shortening Fraction - a heart function test
MM Multiple Myeloma
RBC Red blood cell / red blood count
WBC White blood cell count
WCC White cell count

 

5-FU 5-Fluorouracil (anti cancer drug)
6-MP 6-Mercaptopurine (anti cancer drug)
6-TG 6-Thioguanine (anti cancer drug)
AA Anaplastic Anaemia
ABC Advanced Breast Cancer
ABMT Autologous bone marrow transplant
ADR Adverse Drug Reaction
AE Adverse event
AFP Alphafetoprotein - eg. expressed by germ cell tumours and other cancers
AIDS Acquired immune deficiency syndrome
ALAT Alanine aminotransferase / alinine transaminase
ALCL Anaplastic Large-cell Lymphoma
ALL Acute lymphoblastic leukaemia
ALT Alanine Aminotransferase
AMKL acute megakaryocytic leukemia
AML Acute Myeloid leukaemia
ANC Absolute neutrophil count
ANED Alive no evidence of disease
ANLL Acute non-lymphatic leukaemia
ARMS Alveolar rhabdomyosarcoma
ASR Age Standardised Rate (Incidence)
AUC Area under the curve
B-ALL B-cell Acute Lymphoblastic Leukaemia
BAER Brainstem Auditory Evoked Responce
BCC Basal Cell Carcinoma
BID / BD Twice a day (bis in die)
BM Bone Marrow
BM Blood Monitoring (eg for glucose)
BMJ British Medical Journal
BMR Basal Metabolic Rate
BMT Bone Marrow Transplant
BNF British National Formulary
BP Blood pressure
BRM Biological Response Modifier
BSA Body Surface Area
BSE Breast Self Examination
Bx Biopsy
C/O Complaining of
C/W Continue With
C1 - C7 Cervical vertebrae (spine eg. C7 = seventh cervical vertebra)
Ca Cancer; carcinoma
Ca Calcium
cALL Common ALL
CAT Computerised axial tomography (scan)
cc Cubic centimeter
CCF Congestive Cardiac Failure
CCR Continuous complete remission
CEA Carcinoembryonic Antigen (tumour marker)
CGH Comparative Genomic Hybridisation - cytogenetics method
CGL Chronic Granulocytic Leukaemia
cGy Centi Gray (unit of radiation)
CHF Congestive heart failure
CLL Chronic lymphocytic Leukaemia
cm centimeter - 0.01 meters
CML Chronic myeloid leukaemia
CMML chronic myelomonocytic leukemia
CMV Cytomegalo virus
CNS Central nervous system - the brain and spine
CPM Cyclophosphamide (anti cancer drug)
CPR Cardio pulmonary resuscitation
CR Complete remission / complete response
CRA Clinical Research Associate
CRC Colorectal carcinoma
CRF Case Report Forms
CRF Chronic renal failure
CRO Contract Research Organisation
CSF Cerebro spinal fluid
CSF Colony-stimulating Factor
CT Computerised axial tumography (scan)
CT Chemotherapy
CTC Common Toxicity Criteria
CTCL Cutaneous T-Cell Lymphoma
CTO Clinical Trials Office
CTX Clinical Trials Exemption
CVA Cardiovascular Accident (stroke)
CVC Central venous catheters
CVP Central Venous Pressure
CXR Chest X-Ray
D/C Discharge
D/H Drug History
D/W Discussed With
DCIS Ductal Carcinoma In Situ - type of breast cancer
DDAVP Desmopressin test for urine osmolality
DDx Differential diagnosis
DFI Disease Free Interval
DFS Disease Free Survival - time without disease prior to relapse or last follow-up
DI Diabetes Incipidus
dl deciletre - 0.01 litres
DLBCL Diffuse Large B-cell Lymphoma
DLCL Diffuse large-cell lymphoma
DLS Date last seen
DLT Dose limiting toxicity - determined by phase 1 studies
DMC Data Monitoring Commitee
DNA Deoxyribonucleic acid
DNA Did Not Attend (clinic)
DNR Do Not Resusitate
DOA Dead on Arival
Dx Diagnosis
EBM Evidence-Based Medicine
EBV Epstein-Barr Virus
ECG Electrocardiogram - heart scan
EDTA ethylendiaminetetraacetic acid - used in measuring kidney function
EEG Electroencephalogram - brain scan
EFS Event Free Survival - time from diagnosis to defined events (eg relapse or deat
EJC European Journal of Cancer
EMC Extraskeletal Myxoid Chondrosarcoma
EMUO Early Morning Urine Osmolality (evaluating urine concentration)
ENT Ear nose throat
ESR Erythrocyte Sedimentation rate
ETS Environmental Tobacco Smoke
F/H Family history
FAB French American and British classification scheme for leukaemia
FBC Full Blood Count
FEV Forced expectorant volume (a lung test)
FFA For Further Appointment
FIGO Federation Internat. Gyn. Obst. (FIGO Gynaecological staging system)
FISH Flourescence in situ Hybridisation
FMTC Familial Medullary Thyroid Carcinoma
FNA Fine Needle Aspiration - a type of biposy using a thin needle (or FNAB)
FU Follow up
FVC Forced Vital Capacity
g gram - unit of weight
G-CSF Granulocyte colony stimulating factor promotes production of white blood cells
GA General Anaesthetic
GCP Good Clinical Practice (guidelines)
GCT Germ Cell Tumour
GCT Giant Cell Tumour Context: bone tumours
GFR Gromerular filtration rate
GI Gastrointestinal
GIST Gastrointestinal Stromal Tumours
GM-CSF Granulocyte and macrophage colony stimulating factor
GPR Good Partial Remission
GU Genito-urinary
GvHD Graft versus Host Disease
Gy Grays (units of radiation)
H&E Hematoxylin and Eosin (stain)
H/O History of
Hb Haemoglobin
HCC Hepatocellular Carcinoma
HCG Human Chorionic Gonadotrophin (hormone)
HCL Hairy Cell Leukaemia
HCO3 Bicarbonate
HD Hodgkin's Disease (lymphoma)
HD High dose
HDC High Dose Chemotherapy
HIV Human Immunodeficiency Virus
HL-A Human Leukocyte Associated antigens (HL-A matching for BMT)
HNPCC Hereditary NonPolyposis Colorectal Cancer
HPV Human Papilloma Virus - implicated in some gynacological cancers
HR High risk
HRT Hormone replacement therapy
HTLV Human T-cell leukemia-lymphoma virus
I-131 Radioactive Iodine
ICD International Classification of Diseases (coding system)
ICDO International Classification of Diseases for Oncology (coding system)
ICF Intercellular fluid
ICU Intensive Care Unit
IL2 Interleukin2
IM Intramuscular - into a muscle
IMRT Intensity-Modulated Radiotherapy
INSS International Neuroblastoma Staging System
ITU Intensive Therapy Unit
IU International units
IV Intravenous - into a vein
IVP Intravenous Pyelogram - type of Xray after injection with iodine dye
JCO Journal of Clinical Oncology
K+ Potassium
kg Kilogram - a thousand grams
l liter - unit of volume
L1 - L5 Lumbar vertebrae 1 - 5 (spine eg. L1 = 1st lumbar vertebra)
LCH Langerhans cell histiocytocis
LCIS Lobular Carcinoma In Situ - type of breast cancer
LDH Lactic dehydrogenase -high levels correlate with tumour volume in some cancers
LMM Lentigo Maligna Melanoma
LMP Low Malignant Potential (context: ovarian tumours)
LN Lymph Node
LP Lumbar puncture
LVEF Left Ventricular Fjection Fraction - a heart function test
LVSF Left Ventricular Shortening Fraction - a heart function test
Lx Lumpectomy
m meter (unit of length)
M/H Medical history
MAB - mAb Monoclonal antibody
MDR Multi drug resistant
MDS Myelo dysplastic syndrome
MEN Multiple Endocrine Neoplasia - (familial) a.k.a. FMEN
mEq/l milliequivalent per liter
mets Metastases (where the tumour has spread to secondary sites)
Mg Magnesium
mg milligram - 0.001 gram
MI Miocardial Infarction
mIBG Radioactive Iodine Metaidobenzoguanidine (mIBG scans or mIBG therapy).
ml millilitre 0.001 liter
MM Malignant Melanoma
mM millimole
mm millimeter - 0.001 meters
MM Multiple Myeloma
mOsm milliosmole
MPNST Malignant Peripheral Nerve Sheath Tumour
MPO Medical and Pediatric Oncology (journal)
MRI Magnetic resonance imaging (scan)
MRT Malignant Rhabdoid Tumour
MSSU Mid stream specimen urine
MTD Maximum tolerated dose - phase 1 studies
MTX Methotrexate (anti cancer drug)
MUD Matched Urelated Donor - for bone marrow transplant
Mx Mastectomy
N/V Nausea and vomiting
Na+ Sodium
NAD No Abnormality Detected
NBCCS Nevoid basal cell carcinoma syndrome
NBM Nil by mouth
NED No evidence of disease
ng nanogram - 0.000000001 gram
NHL Non Hodgkin's Lymphoma
NK Natural Killer cells (large lymphocytes, part of the immune system)
NK Not known
NM Nodular Melanoma
NMR Nuclear magnetic resonance (scan)
NMSC Non Melanoma Skin Cancer
NOS Not otherwise specified (see ICDO)
NPC Nasopharyngeal Carcinoma
NRSTS Non-Rhabdomyosarcoma Soft Tissue Sarcoma
NSCLC Non-small cell lung cancer
NSE Neuron-Specific Enolase - a neural marker
NSR Non significant result
NSR Normal Sinus Rhythem
O/E On Examination
ONB Olfactory Neuroblastoma
OS Overall Survival
OS Osteogenic sarcoma (context bone tumours)
PBSC Peripheral Blood Stem Cell (see PBSCT)
PBSCH Peripheral Blood Stem Cell Harvest
PBSCR Peripheral Blood Stem Cell Rescue (transplant)
PBSCT Peripheral Blood Stem Cell Transplant
PD Progressive disease
PDQ Physician's Data Query (CancerNet)
PET Positron Emmission Tomography- a scan after a small radioactive injection.
PET Pancreatic Endocrine Tumor
PFS Progression Free Survival
pg picogram - 0.000000000001 gram
pH hydrogen-ion concentration - acid / alkaline
PH Past History
PLB Primary Lymphoma of Bone
PNET Primitive neuroectodermal tumour Context: CNS tumours
PNET Peripheral neuroectodermal tumour Context: Bone tumours - see Ewing's tu
PNS Peripheral nervous system - nervous system outside the brain and spine.
PR Partial Responce / Partial Remission
PR per rectum
prn as required
prn whenever necessary (pro re nata)
PSA prostate-specific antigen - PSA test used in screening for prostate cancer
PUD Peri-Urethral Diathermy (associated with superficial bladder cancer)
QALY Quality-Adjusted Life Year
qid Four times a day (quater in die)
QoL Quality of Life
RBC Red blood cell / red blood count
RCT Randomised Controlled Trial
RFS Relapse free survival - Time from diagnosis to relapse or death.
RIGS Radioimmunoguided surgery
RMS Rhabdomyosarcoma
RNA ribonucleic acid
RT Radiotherapy
RTPCR Reverse transcriptase polymerase chain reaction
Rx Treatment
SA Surface area (see BSA)
SAE Serious Adverse Event
SC Subcutaneous
SCC Squamous Cell Carcinoma
SCLC Small cell lung cancer
SD Stable Disease
SDV Source Data Verification
SGOT Serum glutamic oxalacetic transaminase - a liver function test
SGPT Serum glutamic pyruvic transaminase - a liver function test
SH Social history
SHO Senior House Officer
SIADH Syndrome of Inappropriate Antidiuretic Hormone
SNP Single Nucleotide Polymorphism
SOB Short of breath
SSM Superficial Spreading Melanoma
T1 - T12 Thoracic vertebrae 1-12 (spine eg. T10 = tenth thoracic vertibra)
TAMs tumour-associated macrophages
TBI Total body irradiation
TCC Transitional Cell Carcinoma (usually bladder cancer)
TCP Thrombocytopenia
tds / tid Three times a day (ter in die)
TNF Tumour Necrosis Factor
TNM Staging system - primary tumour
TPN total parenteral nutrition
TRK Transketolase
U&Es Urea and Electrolites
UA Urine analysis
ug microgram - 0.000001 gram
ULN Upper Limits of Normal
URTI Upper respiratory tract infection
US Ultasound (scan)
UTI Urinary Tract Infection
UVR Ultra Violet Radiation
VEF Ventricular ejection fraction (tests lung function)
VM-26 Teniposide (anti cancer drug)
VMA Vanillylmandelic Acid
VP-16 Etoposide (anti cancer drug)
WBC White blood cell count
WCC White cell count
XRT Radiotherapy (external)
YST Yolk sac tumour - (aka. germ cell tumour)

 

Acronyms of Cancer Organisations

Shortcuts: ABCDEFGHIJKLMNOPQRSTUVWXYZ

 

AACR American Association for Cancer Research
ABPI Association of the British Pharmaceutical Industry
ABTA American Brain Tumour Association
ACDM Association for Clinical Data Management
ACRPI Association of Clinical Research for the Pharmaceutical Industry
ACS American Cancer Society
AIOM Italian Association for Medical Oncology
AIRO Associazione Italiana di Radioterapia Oncologica
AJCC American Joint Committee on Cancer
ASCO American Society of Clinical Oncology
ASH American Society for Hematology
ASTRO American Society for Therapeutic Radiology and Oncology
BACR British Association for Cancer Research
BASO British Association of Surgical Oncologists
BMA British Medical Association
BNLI British National Lymphoma Investigation
BOA British Oncology Association
BODMA British Oncology Data Managers Association (UK)
CALGB Cancer and Leukemia Group B (USA)
CCG Children's Cancer Group (USA)
CCRG Children's Cancer Research Group (Oxford, UK)
CCS Canadian Cancer Society
COG Cooperative Oncology Group (USA)
COS Canadian Oncology Society
CRC Cancer Research Campaign (UK)
CSM Committee on Safety of Medicines (UK)
EACR European Association for Cancer Research
EANO European Association for NeuroOncology
EBMT European Group for Blood and Marrow Transplantation
ECOG Eastern Cooperative Group (USA)
EICESS European Intergroup Ewing's Sarcoma Study
ENSG European Neuroblastoma Study Group
EOI European Osteosarcoma Intergroup
EORTC European Organisation for Research and Treatment of Cancer
ESO European School of Oncology
ESTRO European Society for Therapeutic Radiation and Oncology
FDA Food and Drug Administration (USA)
FECS Federation of European Cancer Societies
GMC General Medical Council (UK)
GPOH Gesellschaft fur Padiatrische Onkologie und Hamatologie (German Paed. Onc Group)
IACR International Association of Cancer Registries
IARC International Agency for Research on Cancer
IASLC International Association for the Study of Lung Cancer
ICARE International Cancer Alliance for Research and Education (ICARE)
ICCCPO International Confederation of Childhood Cancer Parent Organisations
ICCG International Collaborative Cancer Group
ICCPO Icelandic Childhood Cancer Parent Organisation
ICH International Conference on Harmonization (GCP)
ICRF Imperial Cancer Research Fund (UK)
IESS Intergroup Ewing's Sarcoma Study (USA)
IMF International Myeloma Foundation
INFA International Neurofibromatosis Association
IPSO International Society of Pediatric Surgical Oncology
ISNCC International Society of Nurses in Cancer Care
IWMF International Waldenstrom's Maroglubulinemia Foundation
LREC Local Research Ethics Committee (UK)
LRF Leukaemia Research Fund (UK)
LRFA Lymphoma Research Foundation of America
MCA Medicines Control Agency (UK)
MRC Medical Research Council (UK)
MREC Multi-centre Research Ethics Committee (UK)
NAACCR North American Association of Central Cancer Registries
NABCO National Alliance of Breast Cancer Organizations
NCCF National Childhood Cancer Foundation (USA)
NCI National Cancer Institue (USA)
NCIC National Cancer Institute of Canada
NCRA National Cancer Registrars Association (USA)
NECCR North of England Children's Cancer Research Unit
NKCA National Kidney Cancer Association (USA)
NNFF National Neurofibromatosis Foundation (USA)
NORD National Organization for Rare Disorders (USA)
NRCT National Registry of Childhood Tumours (UK) held by the CCRG at Oxford
OECI Organization of European Cancer Institutes
ONS Oncology Nursing Society (USA)
POG Pediatric Oncology Group (USA)
PONF Paediatric Oncology Nurses Forum (UK)
SCTN Scottish Cancer Therapy Network
SEER Surveillance, Epidemiology, and End Results (USA)
SFOP French Paediatric Oncology Scociety
SGDM Study Group on Data Management (EORTC)
SGO Society of Gynecologic Oncologists
SIOP International Society of Paediatric Oncology
SNLG Scottish and Newcastle Lymphoma Group
SPOHNC Support for People with Oral and Head and Neck Cancer
SWOG Southwest Oncology Group (USA)
UICC Union Internationale Contre le Cancer - International Union Against Cancer
UKACR UK Association of Cancer Registeries
UKCCCR UK Coordinating Committee for Cancer Research
UKCCRG UK Children's Cancer Research Group (Oxford)
UKCCSG UK Children's Cancer Study Group (UK)
WHO World Health Organisation
 

 

Function of the cardiovascular system

Blood circulates through a network of vessels throughout the body to provide individual cells with oxygen and nutrients and helps dispose of metabolic wastes. The heart pumps the blood around the blood vessels. Blood is made up of about 45% solids (cells) and 55% fluids (plasma). The plasma is largely water, containing proteins, nutrients, hormones, antibodies, and dissolved waste products.

Functions of blood and circulation:

  • Circulates OXYGEN and removes Carbon Dioxide.
  • Provides cells with NUTRIENTS.
  • Removes the waste products of metabolism to the excretory organs for disposal.
  • Protects the body against disease and infection.
  • Clotting stops bleeding after injury .
  • Transports HORMONES to target cells and organs.
  • Helps regulate body temperature.

 


Blood

General types of blood cells: (each has many different sub-types)

ERYTHROCYTES (red cells) are small red disk shaped cells. They contain HAEMOGLOBIN, which combines with oxygen in the lungs and is then transported to the body's cells. The haemoglobin then returns carbon dioxide waste to the lungs. Erythrocytes are formed in the bone marrow in the knobby ends of bones.

LEUKOCYTES (white cells) help the body fight bacteria and infection. When a tissue is damaged or has an infection the number of leukocytes increases. Leukocytes are formed in the small ends of bones. Leukocytes can be classed as granular or non granular. There are three types of granular leukocytes (eosinophils, neutrophils, and basophils), and three types of non-granular (monocytes, T-cell lymphocytes, and B-cell lymphocytes). See also the lymphatic system.

THROMBOCYTES (platelets) aid the formation of blood CLOTS by releasing various protein substances. When the body is injured thrombocytes disintegrate and cause a chemical reaction with the proteins found in plasma, which eventually create a thread like substance called FIBRIN. The fibrin then "catches" other blood cells which form the clot, preventing further loss of blood and forms the basis of healing.

 


Blood vessels

ARTERIES carry oxygenated blood away from the heart. They are thick hollow tubes which are highly ELASTIC which allows them to DILATE (widen) and constrict (narrow) as blood is forced down them by the heart. Arteries branch and re-branch, becoming smaller until they become small ARTERIOLES which are even more elastic. Arterioles feed oxygenated blood to the capillaries. The AORTA is the largest artery in the body, taking blood from the heart, branching into other arteries that send oxygenated blood to the rest of the body.

CAPILLARIES distribute the nutrients and oxygen to the body's tissues and remove deoxygenated blood and waste. They are extremely thin, the walls are only one cell thick and connect the arterioles with the venules (very small veins).

VENULES (very small veins) merge into VEINS which carry blood back to the heart. The vein walls are similar to arteries but thinner and less elastic. Veins carry deoxygenated blood towards the lungs where oxygen is received via the pulmonary capillaries. The PULMONARY Veins then carries this oxygenated blood back to the heart.

 


The heart

The heart is a hollow muscular organ which beats over 100,000 times a day to pump blood around the body's 60,000 miles of blood vessels. The right side of the heart receives blood and sends it to the lungs to be oxygenated, while the left side receives oxygenated blood from the lungs and sends it out to the tissues of the body. The Heart has three layers; the ENDOCARDIUM (inner layer), the EPICARDIUM (middle layer), and MYOCARDIUM (outer layer). The heart is protected by the PERICARDIUM which is the protective membrane surrounding it.

The heart has FOUR CHAMBERS, in the lower heart the right and left Ventricles, and in the upper heart the right and left Atria. In a normal heart beat the atria contract while the ventricles relax, then the ventricles contract while the atria relax. There are VALVES through which blood passes between ventricle and atrium, these close in such a way that blood does not backwash during the pauses between ventricular contractions. The right and left ventricles are divided by a thick wall (the VENTRICULAR SEPTUM), babies born with "hole in the heart" have a small gap here, which is a problem since oxygenated and deoxygenated can blood mix. The walls of the left ventricle are thicker as it has to pump blood to all the tissues, compared to the right ventricle which only pumps blood as far as the lungs.

 


The spleen

This is a large flat oval organ located below the diaphragm, it's main function is to STORE BLOOD. The size of the spleen can vary, for example it may enlarge when the body is fighting infection also it's size tends to decrease with age. It is a non-vital organ and it is possible to survive after removal of the spleen.

Perinicious anaemia is a Vitamin B12 deficiency resulting in a reduction in number of erythrocytes.

Aplastic anemia is a failure of the bone marrow to produce the enough red blood cells.

Septicaemia - bacterial toxins in blood.

 


Roots, suffixes, and prefixes

component meaning example
CARDIO- heart echocardiogram = sound wave image of the heart.
CYTE- cell thrombocyte = clot forming cell.
HAEM- blood haematoma - a tumour or swelling filled with blood.
THROMB- clot, lump thrombocytopenia = deficiency of thrombocytes in the blood
ETHRO- red ehtrocyte = red blood cell
LEUKO- white leukocyte = white blood cell
SEP, SEPTIV- toxicity due to micro-organisms septicaemia
VAS- vessel / duct cerebrovascular = blood vessels of the cerebrum of the brain.
HYPER- excessive hyperglycaemia = excessive levels of glucose in blood.
HYPO- deficient / below hypoglycaemia = abnormally low glucose blood levels.
-PENIA deficiency neutropenia = low levels of neutrophilic leukocytes.
-EMIA condition of blood anaemia = abnormally low levels of red blood cells.

 


Cancer Focus

 

Overview of Haematological Malignancies
 
The most common haematological malignancy is leukaemia - cancer of the white blood cells. There are many types of leukaemia; Acute types progress rapidly, while Chronic types develop more slowly. Leukaemia is often accompanied by anaemia because the red oxygen carrying cells in the blood are crowded out by the cancerous white cells. There are a number of malignancies and disorders affecting other types of blood cells.
Internet Resources for Leukaemia

 

Acute Lymphoblastic Leukaemia (ALL)
 
Acute lymphoblastic leukaemia (also known as acute lymphocytic leukaemia or ALL) is a disease where too many immature lymphocytes (a type of white blood cell) are found in the blood and bone marrow. Symptoms can include persistent fever, weakness or tiredness, achiness in the bones or joints, or swollen lymph nodes. Adult ALL and its treatment is usually different to childhood ALL. Almost a third of adult patients have a specific chromosome translocation; "Philadelphia Positive" ALL.
Internet Resources for Acute Lymphocytic Leukaemia

 

Acute Myeloid Leukaemia (AML)
 
Acute myeloid leukemia (AML) is a disease in which too many immature granulocytes (a type of white blood cell) are found in the blood and bone marrow. There are a number of subtypes of AML including acute myeloblastic leukemia, acute promyelocytic leukemia, acute monocytic leukemia, acute myelomonocytic leukemia, erythroleukemia, and acute megakaryoblastic leukemia.
Internet Resources for Acute Myeloid Leukaemia

 

Other Types of Leukaemia
 

Chronic Lymphocytic Leukaemia
Chronic Myelogenous Leukaemia
Hairy Cell Leukaemia
Internet Resources for Leukaemia

 

Childhood Leukaemia
 
Childhood leukaemias tend to have different characteristics and treatments compared to adult leukaemias. There is a "childhood peak" of Acute Lymphoblastic Leukaemia, there is a lower proportion of Acute Myeloid Leukaemias compared to adult patients. Clinical prognostic factors include age, White Blood Cell count (WBC) at presentation, and Central Nervous System (CNS) involvement. Infants less than 1 year and adolescents over 10 years of age, WBC greater than 50,000, or CNS involvement are associated with a less favourable prognosis.
Internet Resources for Childhood Leukaemia

 

Other Haematological Malignancies
 
- Myelodysplastic Syndromes
Myelodysplastic syndromes, sometimes called "pre-leukaemia" are a group of diseases in which the bone marrow does not produce enough normal blood cells. Common symptoms are anaemia, bleeding, easy bruisability, and fatigue. These Myelodysplastic syndromes can occur in all age groups but are more common in people aged over 60. Myelodysplastic syndromes may develop spontaneously or be secondary to treatment with chemotherapy / radiotherapy. There is an association with Myelodysplastic syndromes and acute myeloid leukaemia.
- Myeloproliferative Disorders
Myeloproliferative disorders are diseases in which too many blood cells are made by the bone marrow, there are 4 main types of myeloproliferative disorders: chronic myelogenous leukaemia, polycythemia vera, agnogenic myeloid metaplasia, and essential thrombocythemia. Chronic myelogenous leukaemia is where an excess of granulocytes (immature white blood cells) are found in the blood and bone marrow. Polycythemia vera is where red blood cells become too numerous often resulting in a swelling of the spleen. Agnogenic myeloid metaplasia is a condition in which certain blood cells do not mature properly, this may result in a swelling of the spleen and anaemia. Essential thrombocythemia is a disease in which the body produces excessive numbers of platelets (cells in the blood that make it clot) which impedes the normal circulation of blood.
- Aplastic Anaemia
Anaplastic Anemia is not a cancer. AA is a rare disease in which the bone marrow is unable to produce adequate blood cells; leading to pancytopenia (deficiency of all types of blood cells). AA may occur at any age, but there is a peak in adolescence / early adulthood, and again in old age. Slightly more males than females are diagnosed with AA, also the disease is more common in the Far East. Patients successfully treated for aplastic anemia have a higher risk of developing other diseases later in life, including cancer.
- Fanconi Anaemia
Fanconi Anaemia is not a cancer, it is a rare disorder found in children that involves the blood and bone marrow. The symptoms include severe aplastic anemia, hypoplasia of the bone marrow, and patchy discoloration of the skin. Recent research has shown an association between Fanconi anaemia and leukaemia.
- Waldenstrom's Macroglobulinemia
This is a rare malignant condition, involving an excess of beta-lymphocytes (a type of cell in the immune system) which secrete immunoglobulins (a type of antibody). WM usually occurs in people over sixty, but has been detected in younger adults.
Internet Resources for Haematological Malignancies

 

French-American-British (FAB) Classification Scheme
 
Leukaemia can be classified using the French-American-British (FAB) criteria. for cell morphology:

L1 - ALL: small lymphoid cells, regular nuclei
L2 - ALL: large lymphoid cells, irregualr nuclei
L3 - ALL: large homogeneous cells with prominent nucleolus
M1 - Myeloblastic leukemia without maturation
M2 - Myeloblastic leukemia with maturation
M3 - Promyelocytic leukemia
M4 - Myelomonocytic leukemia
M5 - Monocytic leukemia
M6 - Erythroleukemia
M7 - Megakaryoblastic leukemia
M0 - AML with minimal differentiation

 

CNS Prophylaxis
 
Leukaemia can sometimes spread to the spinal cord and brain (Central Nervous System). Intrathecal chemotherapy (injection into the fluid around the spine) may be given to combat or prevent CNS relapse.

 

Blood Counts
 
Blood counts are done to test the number of each of the different kinds of cells in the blood. This may be an aid to diagnosis or done to monitor toxicity after each course of chemotherapy. The next course of chemotherapy may be delayed until white cells, neutrophils, and platelets have recovered to a safe level.

 

Cardiotoxicity
 
Cardiotoxicity (damage to the heart) is associated with certain anti cancer drugs, especially Adriamycin. As such the total dose of these drugs may be limited to reduce the risk of cardiotoxicity.

 

Echocardiagram
 
An Echocardiogramis where an image of the heart is formed when high frequency sound waves are reflected from the muscles of the heart. An echocardiogram may be done before treatment starts to establish a baseline from which to compare future tests.

 

Metastases through the cardivascular system
 
The network of blood vessels reach all parts of the body and may provide one of the routes for cancer cells to spread to secondary sites.

Related Abbreviations and Acronyms:

 

 
AA Anaplastic Anaemia
ALL Acute lymphoblastic leukaemia
AML Acute Myeloid leukaemia
ANC Absolute neutrophil count
ANLL Acute non-lymphatic leukaemia
ASH American Society for Hematology
B-ALL B-cell Acute Lymphoblastic Leukaemia
BP Blood pressure
CALGB Cancer and Leukemia Group B (USA)
cALL Common ALL
CGL Chronic Granulocytic Leukaemia
CHF Congestive heart failure
CLL Chronic lymphocytic Leukaemia
CML Chronic myeloid leukaemia
CMML chronic myelomonocytic leukemia
CPR Cardio pulmonary resuscitation
CVA Cardiovascular Accident (stroke)
CVC Central venous catheters
ECG Electrocardiogram - heart scan
FAB French American and British classification scheme for leukaemia
FBC Full Blood Count
G-CSF Granulocyte colony stimulating factor promotes production of white blood cells
GM-CSF Granulocyte and macrophage colony stimulating factor
Hb Haemoglobin
HCL Hairy Cell Leukaemia
HD Hodgkin's Disease (lymphoma)
HTLV Human T-cell leukemia-lymphoma virus
IV Intravenous - into a vein
LRF Leukaemia Research Fund (UK)
LVEF Left Ventricular Fjection Fraction - a heart function test
LVSF Left Ventricular Shortening Fraction - a heart function test
MM Multiple Myeloma
RBC Red blood cell / red blood count
WBC White blood cell count
WCC White cell count

More Cancer Related Abbreviations